FDA Medical Device Recalls (openFDA)
St Jude Medical Inc.: Merlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model 3650 Programmer is a portable, dedicated programming system designed to interrogate program display data and test St. Jude Medical implantable devices.
- Recall number
- Z-0032-2018
- Recalling firm
- St Jude Medical Inc.
- Product
- Merlin PCS 3650 Programmer with Model 3330 Software Version 23.1.1 The Merlin Patient Care System (PCS) Model 3650 Programmer is a portable, dedicated programming system designed to interrogate program display data and test St. Jude Medical implantable devices.
- Status
- Terminated
- Initiated
- 2017-08-28
- Posted
- not on file
- Terminated
- 2020-07-08
- Reason
- New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- Root cause
- Unknown/Undetermined by firm
- Action
- Abbott sent an Important Cybersecurity Advisory dated August 28, 2017, to all affected customers to notify customers of the availability of the programmer software update and associated pacemaker firmware update. The notification material is in the form of a Physician Letter and Hospital Letter and will be delivered to physicians and hospitals with inventory in the US by overnight service. Customers with questions were instructed to contact their Abbott representative or the customers technical support hotline at 1-800-722-3774.
- Quantity
- 10,453 units
- Distribution
- Worldwide Distribution
- Lot, serial or model codes
- Software version 23.1.1
- 510(k) numbers
- not on file
- PMA numbers
- ["P030054S135","P030054S329"]
- Product code
- NIK
- Firm FEI
- 2017865
- Firm city
- Sylmar
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28