NOX MEDICAL: Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep.
- Recall number
- Z-0031-2022
- Recalling firm
- NOX MEDICAL
- Product
- Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep.
- Status
- Terminated
- Initiated
- 2021-09-07
- Posted
- not on file
- Terminated
- 2023-05-01
- Reason
- Error in the device firmware results in the inaccurate detection of device position.
- Root cause
- Finished device change control
- Action
- Written notification letter titled, "URGENT: Medical Device Recall" was sent to all consignees on 09/07/2021. The letter instructs the consignee to follow the following instructions: "Further distribution or use of the affected devices should cease immediately. Devices should be quarantined pending device firmware update, per the instructions below. Due to the potential of inaccurate position-related reporting of TST, TRT, AHI and REI; studies performed with the affected T3s devices should be reanalyzed to determine if this has clinical impact on the patient diagnoses. A Nox Medical technical representative will contact you within 5 business days to schedule a remote session (via videoconference or other means) to upgrade the device firmware. Nox Medical requests that you formally acknowledge receipt of this notice and return the attached Customer / Distributor - Acknowledgement and Receipt Form (Attachment 1). This recall notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. If any affected devices which were in your possession (indicated in Attachment 2: Product Distribution and Information Table) have been transferred, please indicate this and provide as much detail as possible in the Acknowledgement and Receipt Form (Attachment 1)."
- Quantity
- 128 units
- Distribution
- US Nationwide distribution in the states of OH, IL, GA, IA, CA, PA, NM, MI, SD.
- Lot, serial or model codes
- Catalog Number: 561111; UDI: (01)15694311111443; Serial Numbers: 300106423, 300106426, 300106420, 300106419, 300106422, 300106427, 300106425, 300106424, 300106421, 300106477, 300106485, 300106783, 300106806, 300106841, 300106843 , 300106846, 300106848, 300106852, 300106853, 300106857, 300106862, 300106881, 300106882, 300106883, 300106884, 300106885, 300106886, 300106888, 300106889, 300106890, 300106894, 300106897, 300106912, 300107004, 300106385, 300106396, 300106397, 300106407, 300106413, 300106414, 300106399, 300106415, 300106416, 300106475, 300106476, 300106483, 300106486, 300106494, 300106534, 300106537, 300106538, 300106544, 300106545, 300106547, 300106548, 300106552, 300106648, 300106658, 300106675, 300106677, 300106678, 300106684, 300106754, 300106772, 300106775, 300106807, 300106809, 300106810, 300106812, 300106839, 300106842, 300106849, 300106855, 300106868, 300106891, 300106892, 300106899, 300106909, 300106926, 300106951, 300107003, 300106398, 300106481, 300106542, 300106546, 300106782, 300106910, 300106400, 300106535, 300106536, 300106808, 300106950, 300106469, 300106470, 300106869, 300106854, 300106478, 300106805, 300106811, 300106840, 300106850, 300106856, 300106859, 300106887, 300106893, 300106895, 300106896, 300106898, 300107005, 300106776, 300106781, 300106789, 300106788, 300106833, 300106549, 300106543, 300106557, 300106540, 300106533, 300106556, 300106555, 300106901, 300106786, 300106553, 300106784, 300106541, 300106777, 300106774
- 510(k) numbers
- ["K082113"]
- PMA numbers
- not on file
- Product code
- MNR
- Firm FEI
- 3007389703
- Firm city
- Reykjavik
- Firm state
- not on file
- Firm country
- Iceland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28