Skip to the content

FDA Medical Device Recalls (openFDA)

NOX MEDICAL: Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep.

Recall number
Z-0031-2022
Recalling firm
NOX MEDICAL
Product
Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep.
Status
Terminated
Initiated
2021-09-07
Posted
not on file
Terminated
2023-05-01
Reason
Error in the device firmware results in the inaccurate detection of device position.
Root cause
Finished device change control
Action
Written notification letter titled, "URGENT: Medical Device Recall" was sent to all consignees on 09/07/2021. The letter instructs the consignee to follow the following instructions: "Further distribution or use of the affected devices should cease immediately. Devices should be quarantined pending device firmware update, per the instructions below. Due to the potential of inaccurate position-related reporting of TST, TRT, AHI and REI; studies performed with the affected T3s devices should be reanalyzed to determine if this has clinical impact on the patient diagnoses. A Nox Medical technical representative will contact you within 5 business days to schedule a remote session (via videoconference or other means) to upgrade the device firmware. Nox Medical requests that you formally acknowledge receipt of this notice and return the attached Customer / Distributor - Acknowledgement and Receipt Form (Attachment 1). This recall notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. If any affected devices which were in your possession (indicated in Attachment 2: Product Distribution and Information Table) have been transferred, please indicate this and provide as much detail as possible in the Acknowledgement and Receipt Form (Attachment 1)."
Quantity
128 units
Distribution
US Nationwide distribution in the states of OH, IL, GA, IA, CA, PA, NM, MI, SD.
Lot, serial or model codes
Catalog Number: 561111; UDI: (01)15694311111443; Serial Numbers: 300106423, 300106426, 300106420, 300106419, 300106422, 300106427, 300106425, 300106424, 300106421, 300106477, 300106485, 300106783, 300106806, 300106841, 300106843 , 300106846, 300106848, 300106852, 300106853, 300106857, 300106862, 300106881, 300106882, 300106883, 300106884, 300106885, 300106886, 300106888, 300106889, 300106890, 300106894, 300106897, 300106912, 300107004, 300106385, 300106396, 300106397, 300106407, 300106413, 300106414, 300106399, 300106415, 300106416, 300106475, 300106476, 300106483, 300106486, 300106494, 300106534, 300106537, 300106538, 300106544, 300106545, 300106547, 300106548, 300106552, 300106648, 300106658, 300106675, 300106677, 300106678, 300106684, 300106754, 300106772, 300106775, 300106807, 300106809, 300106810, 300106812, 300106839, 300106842, 300106849, 300106855, 300106868, 300106891, 300106892, 300106899, 300106909, 300106926, 300106951, 300107003, 300106398, 300106481, 300106542, 300106546, 300106782, 300106910, 300106400, 300106535, 300106536, 300106808, 300106950, 300106469, 300106470, 300106869, 300106854, 300106478, 300106805, 300106811, 300106840, 300106850, 300106856, 300106859, 300106887, 300106893, 300106895, 300106896, 300106898, 300107005, 300106776, 300106781, 300106789, 300106788, 300106833, 300106549, 300106543, 300106557, 300106540, 300106533, 300106556, 300106555, 300106901, 300106786, 300106553, 300106784, 300106541, 300106777, 300106774
510(k) numbers
["K082113"]
PMA numbers
not on file
Product code
MNR
Firm FEI
3007389703
Firm city
Reykjavik
Firm state
not on file
Firm country
Iceland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register