Aesculap, Inc.: UNITRAC Pneumatic Retraction Arm. A support arm used to hold surgical instruments during surgical procedures.

Recall number
Z-0031-2014
Recalling firm
Aesculap, Inc.
Product
UNITRAC Pneumatic Retraction Arm. A support arm used to hold surgical instruments during surgical procedures.
Status
Terminated
Initiated
2013-07-16
Posted
2013-10-21
Terminated
2014-08-26
Reason
Changes were implemented to the UNITRAC instructions for use to ensure safe use and potentially extend the life of the product. This included recommending the use of Sterile Cover JG901 and new Quick Release Coupling RT02OR to eliminate the need to sterilize the UNITRAC and a functional check was added prior to use to ensure optimal functionality of the UNITRAC system.
Root cause
Process change control
Action
A letter dated July 16, 2013 was sent requesting users to review the new instructions for use in their entirety, sign the attached Hospital Acknowledgment form and return to Aesculap. An Aesculap representative would be in contact to provide an in-service regarding the process for set-up of the Sterile Cover and required Functional Check. Discontinuation of use and repair of the device was recommended if the joints did not move freely or if the set position drifts during the Functional Check.
Quantity
217
Distribution
Nationwide Distribution.
Lot, serial or model codes
RT040R, all lot numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAD
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register