Baxter Healthcare Corporation: L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide.

Recall number
Z-0030-2020
Recalling firm
Baxter Healthcare Corporation
Product
L5C4531 X1 Integrated APD Set with Cassette 3-Prong Sterile, nonpyrogenic fluid path. Sterilized by ethylene oxide.
Status
Terminated
Initiated
2018-10-10
Posted
not on file
Terminated
2020-07-22
Reason
Potential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may predispose the patient to peritonitis.
Root cause
Mixed-up of materials/components
Action
On 10/10/2018, and 10/15/2018, Urgent Product Recall notices were mailed to dialysis providers at medical centers. Customers were asked to do the following: 1. Locate and remove the affected product lot from your facility. The product code and lot number can be found on the individual product or shipping carton. 2. Contact recalling firm for Service to arrange for return and credit. Center for Service can be reached at 888-229-0001. 3. If you purchased this product directly from the recalling firm, complete and return the enclosed customer reply form. 4. If you purchased this product from a distributor, please contact your distributor for return and credit. If a reply form is provided by your distributor or wholesaler, please return it to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a consumer-level recall of the affected product that you distributed to customers. Customers with additional questions are encouraged to contact: Corporate Product Surveillance at 800-437-5176, Clinical Helpline at 888-736-2543, and Renal Therapy Services at 800-553-6898. On 10/15/2018, Urgent Product Recall notices were mailed to Peritoneal Dialysis Patients who were asked to do the following: 1. Locate and remove the affected product lot from your supplies. The product code and lot number can be found on the individual product or shipping carton. 2. Contact recalling firm to arrange for return and credit. HomeCare Services can be reached at 800-284-4060. 3. Complete and return the enclosed Home Patient Reply Form.
Quantity
8640 eaches
Distribution
U.S.: VA, DC, CA, FL, SC, MD, DE, PA, NC, WA, GA, TX, NV. No foreign (OUS) distribution.
Lot, serial or model codes
Lot Number H18D22030
510(k) numbers
["K102936"]
PMA numbers
not on file
Product code
FKX
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register