Tcoag Us, Incorporated: Minutex D-Dimer ; Trinity Biotech plc, Bray, Co. Wicklow, Ireland. Tel: (353) 1 276 9800, Fax: (353) 1 276 9888, Web: www.trinitybiotech.com Minutex D-Dimer is a latex agglutination test for semiquantitative determination of fibrin D-Dimer.

Recall number
Z-0030-2012
Recalling firm
Tcoag Us, Incorporated
Product
Minutex D-Dimer ; Trinity Biotech plc, Bray, Co. Wicklow, Ireland. Tel: (353) 1 276 9800, Fax: (353) 1 276 9888, Web: www.trinitybiotech.com Minutex D-Dimer is a latex agglutination test for semiquantitative determination of fibrin D-Dimer.
Status
Terminated
Initiated
2011-06-13
Posted
2011-10-07
Terminated
2012-12-11
Reason
There is a performance issue with Minutex D-Dimer Kit which may cause false negative results for patient samples containing levels of greater than 250ng/mL D-Dimer.
Root cause
Nonconforming Material/Component
Action
The firm,Tcoag US, Incorporated, sent an "Urgent Device Recall" letter dated June 13, 2011 via UPS to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately examine their inventory and quarantine the products subject to recall; complete and return the attached Product Recall Notification Response Form by June 22, 2011 via fax at 973-775-3085; scrap the products consistent with their laboratory policy and if they have further distributed these products, notify their customers to quarantine and scrap the recalled product. For questions or technical assistance, please call the Tcoag hotline at 1-888-291-0415, option 2 (7.30 am to 6.00 pm Eastern Time).
Quantity
4409 Units
Distribution
Worldwide distribution: USA (nationwide) and country of: Canada.
Lot, serial or model codes
510 K number: K920668  Device listing number: D099358 Lot Numbers: Expiry Date: C040002 10 July 2012 C042002 14 Jan 2012 B218005 05 Jan 2012 B341002 03 May 2012 B144001 30 Sept 2011
510(k) numbers
["K920668"]
PMA numbers
not on file
Product code
GHH
Firm FEI
3005261996
Firm city
Budd Lake
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register