Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Corp.: Roche/Hitachi brand lactate reagent for the quantitative determination of L-lactate in plasma, cerebrospinal fluid or whole blood. Catalog # 1822837.

Recall number
Z-0028-04
Recalling firm
Roche Diagnostics Corp.
Product
Roche/Hitachi brand lactate reagent for the quantitative determination of L-lactate in plasma, cerebrospinal fluid or whole blood. Catalog # 1822837.
Status
Terminated
Initiated
2003-09-02
Posted
2003-10-15
Terminated
2003-12-10
Reason
Patient samples with very high lactate values may falsely be reported as within the normal range without the instrument giving a warning flag to the technician.
Root cause
Other
Action
Recall letters were sent to U.S. customers on 9/2/03 instructing them not to run lactate with these analyzers. An update notification letter was sent on or about 9/18/03 telling customers to discontinue use of the product for manual lactate determinations.
Quantity
21958
Distribution
United States, Canada and Germany
Lot, serial or model codes
All lots of reagent when used manually or on the Roche/Hitachi 902, 911, 912, 917 or Modular Analytics P module analyzers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LNN
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register