Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016

Recall number
Z-0027-2024
Recalling firm
Bard Peripheral Vascular Inc
Product
Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016
Status
Open, Classified
Initiated
2023-08-10
Posted
2023-10-04
Terminated
not on file
Reason
Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial cannula and accessing target tissue, which may lead to repeated procedure, and procedural complications, such as pain/discomfort, pneumothorax, and bleeding.
Root cause
Device Design
Action
On 8/10/23, BD mailed or emailed recall notices to Medical Directors, Risk Managers, Medical Device Safety Offices, Lab Managers, Clinical Engineers, and Distributors who were told to do the following: 1) Discontinue distribution and use and immediately dispose of all affected product. 2) Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. 3) Distributors, identify all customers within your distribution network that purchased any affected product and provide a copy of the customer letter to all customers and advise them of this field action notification on behalf of the recalling firm. 2) Complete the attached Response Form and return via email to: BDRC45@bd.com If you require further assistance contact: the North American Regional Complaint Center: 1-844-8BD- LIFE (1-844-823-5433) M-F 8am - 5pm CT, or Email: productcomplaints@bd.com
Quantity
1,545
Distribution
Worldwide - US Nationwide distribution in the states of PR, CO, TX, CA, WA, OR, AZ, NJ, NY, OH, NC, MD, MI, LA, IN, AK, TN and the countries of CA, AU, BE.
Lot, serial or model codes
UDI-DI: 00801741097393, Lot: 0001480774, Exp: 07/28/2025
510(k) numbers
["K133948"]
PMA numbers
not on file
Product code
KNW
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register