Hologic, Inc: Panther(R) System, Catalog Number 902615, For Use with Hologic Assays, For In Vitro Diagnostic Use

Recall number
Z-0027-2018
Recalling firm
Hologic, Inc
Product
Panther(R) System, Catalog Number 902615, For Use with Hologic Assays, For In Vitro Diagnostic Use
Status
Terminated
Initiated
2017-09-15
Posted
not on file
Terminated
2020-07-30
Reason
An incorrect incubator was installed on one Panther System which potentially led to incorrect Aptima Combo 2(R) assay (AC2) results.
Root cause
Employee error
Action
The customer was notifed of the potential incorrect results via a phone call on September 15, 2017. A customer letter was sent on September 18, 2017. The notification letter requests the customer to re-test any negative Aptima Combo 2 assay results tested between July 13, 2017 Aug 30, 2017.
Quantity
1 unit
Distribution
Indiana
Lot, serial or model codes
Serial Number 20900001124
510(k) numbers
["K132251"]
PMA numbers
not on file
Product code
NSU
Firm FEI
2024800
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register