FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices.
- Recall number
- Z-0027-2016
- Recalling firm
- Zimmer, Inc.
- Product
- Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices.
- Status
- Terminated
- Initiated
- 2015-08-11
- Posted
- 2015-10-06
- Terminated
- 2016-03-15
- Reason
- It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat
- Root cause
- Process control
- Action
- On 8/11/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 a.m. and 5:00 p.m. EST.
- Quantity
- 18 units
- Distribution
- Distributed in NC, AL, KS, TX, UT, CA, and SINGAPORE.
- Lot, serial or model codes
- Item No. 00-4836-080-00; Lot 56555128
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FSM
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as