Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices.

Recall number
Z-0027-2016
Recalling firm
Zimmer, Inc.
Product
Zimmer Universal Locking System, Titanium 3.5mm. Zimmer Implant and Instrument Cases are reusable, non-implantable devices designed to facilitate organization, identification, storage, transportation, cleaning, and sterilization processing of reusable instruments and certain unused implantable devices.
Status
Terminated
Initiated
2015-08-11
Posted
2015-10-06
Terminated
2016-03-15
Reason
It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat
Root cause
Process control
Action
On 8/11/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 a.m. and 5:00 p.m. EST.
Quantity
18 units
Distribution
Distributed in NC, AL, KS, TX, UT, CA, and SINGAPORE.
Lot, serial or model codes
Item No. 00-4836-080-00; Lot 56555128
510(k) numbers
not on file
PMA numbers
not on file
Product code
FSM
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register