Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson Infusion Therapy Systems Inc.: REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO

Recall number
Z-0025-2024
Recalling firm
Becton Dickinson Infusion Therapy Systems Inc.
Product
REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO
Status
Open, Classified
Initiated
2023-08-29
Posted
2023-10-03
Terminated
not on file
Reason
There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
Root cause
Process change control
Action
On August 29, 2023, Becton Dickinson issued an "Urgent Medical Device Recall" notification via. FedEx. Becton Dickinson asked consignees to take the following actions: 1. Immediately review your inventory for Catalog Number 386862, Lot number 2238464. Destroy all unused product subject to the recall following your institution's process for destruction. 2. For indwelling catheters, no further action is needed. 3. If a defective product was used on a patient and the used needle was disposed of without the clinician suffering a needlestick injury, no further action is necessary to further protect the patient or the clinician. 4. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch, Phone: 1-800-FDA 1088 (1-800-332-1088), Mail: MedWatch, HF-2, FDA, 5600 Fisher.s Lane, Rockville, MD 20852-978
Quantity
98,000 units
Distribution
US Nationwide distribution in the states of CA, FL, GA, MA, MN, MO, MS, NJ, OH, TX.
Lot, serial or model codes
Lot # 2238464/ UDI-DI: (01)00382903868629
510(k) numbers
["K220584"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
1710034
Firm city
Sandy
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register