Medical Device Technologies, Inc.: Tru-Core II Disposable Automatic Biopsy Instrument and Tru-Core II URO Disposable Automatic Biopsy Instrument. The Tru-Core Disposable Automatic Biopsy Instrument is used to obtain multiple core samples from soft tissue such as liver, kidney, prostate, breast, ect. The recalling lot's for the product are as follows: Model Number: 763114100X Lot #'s: 92661WY5, 93211XXV, 00141YVE, 00541ZKX, 010920HL, 01522192, 92791X5I, 93271Y1W, 00201YYE, 00621ZQ3, 011320LH, 015221BD, 92791X7X, 93411Y9O, 00211Z0C, 00671ZRN, 011720Q, 015521CL, 92861XCT, 93421YBJ, 00251Z1T, 00761ZZP, 012420U0, 016221H2, 93001XLG, 93431YBI, 00351Z86, 0081201Y, 0138211F, 016721KJ, 93081XPC, 93571YJO, 00461ZE2, 0088206B, 0141215R, 93141XST, 00061YOV, 00541ZK0, 009620BO, 0146219M. Model Number: 763114160X Lot #'s 92661WY7, 00111YPZ, 00671ZRO, 013220Z7, 016221H7, 93131XLS, 00331Z61, 0110201Y, 014125Q, 00882065, 00391Z8Y, 013120YC, 015521CJ. Model Number: 763114200X Lot #'s 92661WY0, 00751ZYJ, 012420U1 Model Number: 763116100X Lot #'s 92791X80, 93491YG9, 00331Z6M, 0083203O, 012420U3, 01672KL, 92791X81, 93501YGA, 00351Z8D, 009620BL, 013120Y9, 92931XDD, 00121YRT, 00431ZE1, 009920D3, 015221AX, 93001XLN, 00131YUG, 00541ZKV, 011020IU, 015221BE, 93081XPH, 00541ZJY, 00601ZOT, 011320LG, 015521CK, 93291Y43, 00251Z1R, 0081201W, 012020QC, 016221H3. Model Number: 763116160X Lot #'s 92661WY3, 00882068, 00351Z8A, 00761ZZQ, 012020QB, 016721KK, 92791X89, 00912084, 00471ZG0, 010320FM, 012420TZ, 92861XD4, 01472193, 00541ZK1, 010920HK, 0141215U, 93001XLR, 00151YVC, 00601ZOR, 0110201V, 015521CM, 93291Y40, 00331Z6J, 00701ZUX, 011720OP, 016221HS Model Number: 763116200X Lot #'s 121KN Model Number: 763116200X Lot #'s 92671X09, 00471ZG1, 00882067, 010320FN, 00211Z09, 00601ZOP, 009920D0, 011020IT Model Number: 763118100X Lot #'s 92661WY8, 93491YG7, 00331Z6L, 00621ZQ2, 009920D2, 0138213F, 92791X7Z, 93491YG8, 00351Z88, 00701ZUW, 010320FL, 01472194, 92861XDS, 00061YOT, 00471ZG2, 00761ZZO, 0110201W, 016221H4, 93091XPL, 00131YUF, 00541ZJZ, 0083203P, 012020QE, 93211XXW, 00251Z1S, 00541ZKU, 009620BM, 012420TX. Model Number: 763118160X Lot #s 92661WY6, 93271Y1V, 00201YYD, 00471ZG3, 012420U2, 016221GZ, 92671WZJ, 93411Y9P, 00251Z1Q, 00671ZRK, 013120YB, 92671WZK, 93421YBK, 00331Z6N, 00701ZUY, 013220Z9, 92861XD0, 00061YOW, 00351Z8E, 00882069, 015221BB, 93001XLM, 00111YQO, 00431ZE3, 009920D4, 015521CO. Model Number: 763118200X Lot #s 92861XCD, 93571YJN, 00331Z6K, 00671ZRM, 011320LF, 01382134, 93001XL6, 0061YOU, 00351Z89,00701ZUZ, 011720OO, 0141215S, 93001XLJ, 00912082, 00391Z8Z, 009620BQ, 012020QA, 01472195, 93081XPJ, 00111YQM, 00471ZG5, 009920D5, 013120Y8, 016221H1, 93211XXX, 00131YUD, 00481ZG4, 010320FJ, 013220Z6, 93400Y9K, 00201YYC, 00621ZQ1, 010920HJ, 0137211I. Model Number: 763120100X Lot #s 93001XLV, 00351Z8B, 0088206A, 01472196, 00331Z60, 00601ZOS, 01202QF. Model Number: 763120160X Lot #s 92661WY1, 00882066, 00111YQP, 00351Z8C, 013120YD, 01472197, 92861XDG, 00912083, 00211Z08, 00621ZQ5, 013220Z8. Model Number: 763120200X Lot #s 00351Z85, 012020QG, 012420TW. Model Number: 763418200X Lot #s 92671WZV, 93291Y41, 00351Z87, 00601ZNH, 010320FK, 0141215T, 92671WZW, 00061YOX, 00431ZE0, 0083203N, 0110201S, 015221BC, 92861XDH, 00131YUE, 00491ZIO, 009620BN, 0011720ON, 015521CN, 93081XPR, 00141YVB, 00541ZKT, 009920D1, 0138213E, 016221H6. Model Number: 763418250X Lot #s 92751X60, 00671ZRL, Model Number: 763418250X Lot #s 92661WY4, 93271Y1N, 00141YVD, 00601ZOQ, 0110201X, 013220ZA, 92751X6P, 93291Y42, 00211Z0B, 00761ZZR, 012020QD, 0137211G, 92781X5G, 93631YJP, 00221Z13, 0081201X, 012420TY, 01382133, 92861XDN, 00912085, 00431ZDZ, 009620BP, 013120YA, 016221H0.

Recall number
Z-0025-2011
Recalling firm
Medical Device Technologies, Inc.
Product
Tru-Core II Disposable Automatic Biopsy Instrument and Tru-Core II URO Disposable Automatic Biopsy Instrument. The Tru-Core Disposable Automatic Biopsy Instrument is used to obtain multiple core samples from soft tissue such as liver, kidney, prostate, breast, ect. The recalling lot's for the product are as follows: Model Number: 763114100X Lot #'s: 92661WY5, 93211XXV, 00141YVE, 00541ZKX, 010920HL, 01522192, 92791X5I, 93271Y1W, 00201YYE, 00621ZQ3, 011320LH, 015221BD, 92791X7X, 93411Y9O, 00211Z0C, 00671ZRN, 011720Q, 015521CL, 92861XCT, 93421YBJ, 00251Z1T, 00761ZZP, 012420U0, 016221H2, 93001XLG, 93431YBI, 00351Z86, 0081201Y, 0138211F, 016721KJ, 93081XPC, 93571YJO, 00461ZE2, 0088206B, 0141215R, 93141XST, 00061YOV, 00541ZK0, 009620BO, 0146219M. Model Number: 763114160X Lot #'s 92661WY7, 00111YPZ, 00671ZRO, 013220Z7, 016221H7, 93131XLS, 00331Z61, 0110201Y, 014125Q, 00882065, 00391Z8Y, 013120YC, 015521CJ. Model Number: 763114200X Lot #'s 92661WY0, 00751ZYJ, 012420U1 Model Number: 763116100X Lot #'s 92791X80, 93491YG9, 00331Z6M, 0083203O, 012420U3, 01672KL, 92791X81, 93501YGA, 00351Z8D, 009620BL, 013120Y9, 92931XDD, 00121YRT, 00431ZE1, 009920D3, 015221AX, 93001XLN, 00131YUG, 00541ZKV, 011020IU, 015221BE, 93081XPH, 00541ZJY, 00601ZOT, 011320LG, 015521CK, 93291Y43, 00251Z1R, 0081201W, 012020QC, 016221H3. Model Number: 763116160X Lot #'s 92661WY3, 00882068, 00351Z8A, 00761ZZQ, 012020QB, 016721KK, 92791X89, 00912084, 00471ZG0, 010320FM, 012420TZ, 92861XD4, 01472193, 00541ZK1, 010920HK, 0141215U, 93001XLR, 00151YVC, 00601ZOR, 0110201V, 015521CM, 93291Y40, 00331Z6J, 00701ZUX, 011720OP, 016221HS Model Number: 763116200X Lot #'s 121KN Model Number: 763116200X Lot #'s 92671X09, 00471ZG1, 00882067, 010320FN, 00211Z09, 00601ZOP, 009920D0, 011020IT Model Number: 763118100X Lot #'s 92661WY8, 93491YG7, 00331Z6L, 00621ZQ2, 009920D2, 0138213F, 92791X7Z, 93491YG8, 00351Z88, 00701ZUW, 010320FL, 01472194, 92861XDS, 00061YOT, 00471ZG2, 00761ZZO, 0110201W, 016221H4, 93091XPL, 00131YUF, 00541ZJZ, 0083203P, 012020QE, 93211XXW, 00251Z1S, 00541ZKU, 009620BM, 012420TX. Model Number: 763118160X Lot #s 92661WY6, 93271Y1V, 00201YYD, 00471ZG3, 012420U2, 016221GZ, 92671WZJ, 93411Y9P, 00251Z1Q, 00671ZRK, 013120YB, 92671WZK, 93421YBK, 00331Z6N, 00701ZUY, 013220Z9, 92861XD0, 00061YOW, 00351Z8E, 00882069, 015221BB, 93001XLM, 00111YQO, 00431ZE3, 009920D4, 015521CO. Model Number: 763118200X Lot #s 92861XCD, 93571YJN, 00331Z6K, 00671ZRM, 011320LF, 01382134, 93001XL6, 0061YOU, 00351Z89,00701ZUZ, 011720OO, 0141215S, 93001XLJ, 00912082, 00391Z8Z, 009620BQ, 012020QA, 01472195, 93081XPJ, 00111YQM, 00471ZG5, 009920D5, 013120Y8, 016221H1, 93211XXX, 00131YUD, 00481ZG4, 010320FJ, 013220Z6, 93400Y9K, 00201YYC, 00621ZQ1, 010920HJ, 0137211I. Model Number: 763120100X Lot #s 93001XLV, 00351Z8B, 0088206A, 01472196, 00331Z60, 00601ZOS, 01202QF. Model Number: 763120160X Lot #s 92661WY1, 00882066, 00111YQP, 00351Z8C, 013120YD, 01472197, 92861XDG, 00912083, 00211Z08, 00621ZQ5, 013220Z8. Model Number: 763120200X Lot #s 00351Z85, 012020QG, 012420TW. Model Number: 763418200X Lot #s 92671WZV, 93291Y41, 00351Z87, 00601ZNH, 010320FK, 0141215T, 92671WZW, 00061YOX, 00431ZE0, 0083203N, 0110201S, 015221BC, 92861XDH, 00131YUE, 00491ZIO, 009620BN, 0011720ON, 015521CN, 93081XPR, 00141YVB, 00541ZKT, 009920D1, 0138213E, 016221H6. Model Number: 763418250X Lot #s 92751X60, 00671ZRL, Model Number: 763418250X Lot #s 92661WY4, 93271Y1N, 00141YVD, 00601ZOQ, 0110201X, 013220ZA, 92751X6P, 93291Y42, 00211Z0B, 00761ZZR, 012020QD, 0137211G, 92781X5G, 93631YJP, 00221Z13, 0081201X, 012420TY, 01382133, 92861XDN, 00912085, 00431ZDZ, 009620BP, 013120YA, 016221H0.
Status
Terminated
Initiated
2010-07-02
Posted
2010-10-08
Terminated
2011-06-28
Reason
Medical Device Technologies, Inc., dba Angiotech, Gaineville, Florida is recalling Tru-Core II Disposable Biopsy Devices due to compromised packaging integrity. The Tru-Core Disposable Automatic Biopsy Instrument is used to obtain multiple core samples from soft tissue such as liver, kidney, prostate, breast, ect. The recall is for the following Model Numbers: 763114100X, 763114160X, 763114
Root cause
Other
Action
Angiotech sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated July 2, 2010 and response card to all customers including distributors. The letter described the product, problem and action to be taken by the customers. The customers were instructed to review their current inventory and segregate the affected lot numbers; arrange for return and replacement of the affected product, and complete and return the Customer Acknowledgment Form via fax to the Quality Assurance department at 1-352-338-0662 or 1-800-333-0440. If you have any questions concerning this notification, contact Shannon Brooks at (352) 338-0440 ext 355 or Katrenia Williams at ext 353.
Quantity
11,123 boxes (111,230 units)
Distribution
Worldwide distribution: USA and countries including: Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Israel, Japan, New Zealand, Peru, South Africa, , Uruguay, and Venezuela.
Lot, serial or model codes
The recalling lot's for the product are as follows:  Model Number: 763114100X Lot #'s: 92661WY5, 93211XXV, 00141YVE, 00541ZKX, 010920HL, 01522192, 92791X5I, 93271Y1W, 00201YYE, 00621ZQ3, 011320LH, 015221BD, 92791X7X, 93411Y9O, 00211Z0C, 00671ZRN, 011720Q, 015521CL, 92861XCT, 93421YBJ, 00251Z1T, 00761ZZP, 012420U0, 016221H2, 93001XLG, 93431YBI, 00351Z86, 0081201Y, 0138211F, 016721KJ, 93081XPC, 93571YJO, 00461ZE2, 0088206B, 0141215R, 93141XST, 00061YOV, 00541ZK0, 009620BO, 0146219M.  Model Number: 763114160X Lot #'s 92661WY7, 00111YPZ, 00671ZRO, 013220Z7, 016221H7, 93131XLS, 00331Z61, 0110201Y, 014125Q, 00882065, 00391Z8Y, 013120YC, 015521CJ.  Model Number: 763114200X Lot #'s 92661WY0, 00751ZYJ, 012420U1  Model Number: 763116100X Lot #'s 92791X80, 93491YG9, 00331Z6M, 0083203O, 012420U3, 01672KL, 92791X81, 93501YGA, 00351Z8D, 009620BL, 013120Y9, 92931XDD, 00121YRT, 00431ZE1, 009920D3, 015221AX, 93001XLN, 00131YUG, 00541ZKV, 011020IU, 015221BE, 93081XPH, 00541ZJY, 00601ZOT, 011320LG, 015521CK, 93291Y43, 00251Z1R, 0081201W, 012020QC, 016221H3.  Model Number: 763116160X Lot #'s 92661WY3, 00882068, 00351Z8A, 00761ZZQ, 012020QB, 016721KK, 92791X89, 00912084, 00471ZG0, 010320FM, 012420TZ, 92861XD4, 01472193, 00541ZK1, 010920HK, 0141215U, 93001XLR, 00151YVC, 00601ZOR, 0110201V, 015521CM, 93291Y40, 00331Z6J, 00701ZUX, 011720OP, 016221HS  Model Number: 763116200X Lot #'s 121KN  Model Number: 763116200X Lot #'s 92671X09, 00471ZG1, 00882067, 010320FN, 00211Z09, 00601ZOP, 009920D0, 011020IT    Model Number: 763118100X Lot #'s 92661WY8, 93491YG7, 00331Z6L, 00621ZQ2, 009920D2, 0138213F, 92791X7Z, 93491YG8, 00351Z88, 00701ZUW, 010320FL, 01472194, 92861XDS, 00061YOT, 00471ZG2, 00761ZZO, 0110201W, 016221H4, 93091XPL, 00131YUF, 00541ZJZ, 0083203P, 012020QE, 93211XXW, 00251Z1S, 00541ZKU, 009620BM, 012420TX.    Model Number: 763118160X Lot #s 92661WY6, 93271Y1V, 00201YYD, 00471ZG3, 012420U2, 016221GZ, 92671WZJ, 93411Y9P, 00251Z1Q, 00671ZRK, 013120YB, 92671WZK, 93421YBK, 00331Z6N, 00701ZUY, 013220Z9, 92861XD0, 00061YOW, 00351Z8E, 00882069, 015221BB, 93001XLM, 00111YQO, 00431ZE3, 009920D4, 015521CO.   Model Number: 763118200X Lot #s 92861XCD, 93571YJN, 00331Z6K, 00671ZRM, 011320LF, 01382134, 93001XL6, 0061YOU, 00351Z89,00701ZUZ, 011720OO, 0141215S, 93001XLJ, 00912082, 00391Z8Z, 009620BQ, 012020QA, 01472195, 93081XPJ, 00111YQM, 00471ZG5, 009920D5, 013120Y8, 016221H1, 93211XXX, 00131YUD, 00481ZG4, 010320FJ, 013220Z6, 93400Y9K, 00201YYC, 00621ZQ1, 010920HJ, 0137211I.   Model Number: 763120100X Lot #s 93001XLV, 00351Z8B, 0088206A, 01472196, 00331Z60, 00601ZOS, 01202QF.   Model Number: 763120160X Lot #s 92661WY1, 00882066, 00111YQP, 00351Z8C, 013120YD, 01472197, 92861XDG, 00912083, 00211Z08, 00621ZQ5, 013220Z8.   Model Number: 763120200X Lot #s 00351Z85, 012020QG, 012420TW.  Model Number: 763418200X Lot #s 92671WZV, 93291Y41, 00351Z87, 00601ZNH, 010320FK, 0141215T, 92671WZW, 00061YOX, 00431ZE0, 0083203N, 0110201S, 015221BC, 92861XDH, 00131YUE, 00491ZIO, 009620BN, 0011720ON, 015521CN, 93081XPR, 00141YVB, 00541ZKT, 009920D1, 0138213E, 016221H6.  Model Number: 763418250X Lot #s 92751X60, 00671ZRL,   Model Number: 763418250X Lot #s 92661WY4, 93271Y1N, 00141YVD, 00601ZOQ, 0110201X, 013220ZA, 92751X6P, 93291Y42, 00211Z0B, 00761ZZR, 012020QD, 0137211G, 92781X5G, 93631YJP, 00221Z13, 0081201X, 012420TY, 01382133, 92861XDN, 00912085, 00431ZDZ, 009620BP, 013120YA, 016221H0.
510(k) numbers
["K982960"]
PMA numbers
not on file
Product code
MJG
Firm FEI
1036710
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register