Skip to the content

FDA Medical Device Recalls (openFDA)

Biomedix, Inc.: Biomedix SELEC-3 I.V. Administration Set. Individually sealed into a Tyvek and LDPE/PET pouch. Secondary packaging includes 48 pouched units per box/case. STERILE

Recall number
Z-0023-2015
Recalling firm
Biomedix, Inc.
Product
Biomedix SELEC-3 I.V. Administration Set. Individually sealed into a Tyvek and LDPE/PET pouch. Secondary packaging includes 48 pouched units per box/case. STERILE
Status
Terminated
Initiated
2014-08-14
Posted
2014-10-06
Terminated
2014-11-13
Reason
Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix performed extensive investigation and found insufficient application of glue. Recall was initiated due to the rate of the failure.
Root cause
Process control
Action
Biomedix sent an PRODUCT RECALL NOTIFICATION letter dated August 14, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please contact us with the quantity remaining in inventory or available to be returned to our facility. All units returned to Biomedix will be replaced with the same SELEC-3 item or a similar design at no charge. Please call (1-800-627-2765 ) for further questions.
Quantity
2,928
Distribution
US Distribution including the states of OH, NC, and TX
Lot, serial or model codes
Model/Product Number: B30-102  Lot #416130  Exp. 12/2016
510(k) numbers
["K142097"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1000115518
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register