Biomedix, Inc.: Biomedix SELEC-3 I.V. Administration Set. Individually sealed into a Tyvek and LDPE/PET pouch. Secondary packaging includes 48 pouched units per box/case. STERILE
- Recall number
- Z-0023-2015
- Recalling firm
- Biomedix, Inc.
- Product
- Biomedix SELEC-3 I.V. Administration Set. Individually sealed into a Tyvek and LDPE/PET pouch. Secondary packaging includes 48 pouched units per box/case. STERILE
- Status
- Terminated
- Initiated
- 2014-08-14
- Posted
- 2014-10-06
- Terminated
- 2014-11-13
- Reason
- Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix performed extensive investigation and found insufficient application of glue. Recall was initiated due to the rate of the failure.
- Root cause
- Process control
- Action
- Biomedix sent an PRODUCT RECALL NOTIFICATION letter dated August 14, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please contact us with the quantity remaining in inventory or available to be returned to our facility. All units returned to Biomedix will be replaced with the same SELEC-3 item or a similar design at no charge. Please call (1-800-627-2765 ) for further questions.
- Quantity
- 2,928
- Distribution
- US Distribution including the states of OH, NC, and TX
- Lot, serial or model codes
- Model/Product Number: B30-102 Lot #416130 Exp. 12/2016
- 510(k) numbers
- ["K142097"]
- PMA numbers
- not on file
- Product code
- FPA
- Firm FEI
- 1000115518
- Firm city
- Bloomington
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28