Skip to the content

FDA Medical Device Recalls (openFDA)

Biomerieux Inc: PREVI Isola System, Catalog numbers 29500 and 29500R, bioMerieux, Inc., Durham, NC. Intended to perform the inoculation and streaking of Prepoured Media (PPM) with liquid microbiological samples.

Recall number
Z-0023-2013
Recalling firm
Biomerieux Inc
Product
PREVI Isola System, Catalog numbers 29500 and 29500R, bioMerieux, Inc., Durham, NC. Intended to perform the inoculation and streaking of Prepoured Media (PPM) with liquid microbiological samples.
Status
Terminated
Initiated
2012-08-28
Posted
2012-10-09
Terminated
2014-09-04
Reason
The instrument may not dispense the specimen onto the pre-poured media resulting in a false negative result.
Root cause
Labeling design
Action
bioMerieux sent an "URGENT PRODUCT CORRECTION NOTICE" letter dated August 28, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. An Acknowledgement Form was included for customers to complete and return. Contact Customer Service at 1-800-682-2666 for questions regarding this notice.
Quantity
350 systems
Distribution
Worldwide Distribution-USA (nationwide)-including DC and the states of CA, CO, CT, FL, GA, IL, IN, KY, LA, MO, NC, NJ, NY, OH, OR, PA, RI, SC, TX, and WI, and the country of Canada.
Lot, serial or model codes
Serial numbers AS180-00001 to AS180-00376.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXG
Firm FEI
1950204
Firm city
Hazelwood
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register