Bio-Rad Laboratories, Inc.: VARIANT B-thalassemia Short Program Reorder Pack, 250 Tests, For the separation and determination os HbA and HbF and as an aid in the identification of abnormal hemoglobins in whole blood. IVD, Made in United States. Contents: 1. Elution Buffer 1, 2 x 1900 mL sodium phosphate buffer 2. Elution Buffer 2, 1 x 1800 mL sodium phosphate buffer 3. Wash/Diluent solution, 1 x 1800 mL deionized water 4. Analytical cartridge, 1 cation exchange cartridge 5. 1 Calibrator/Diluent set, 1 Lypholized human red blood cells hemolysate with preservative; deionized water with preservative 6. 1 CD Rom - CD R with program parameters 7 Sample vials, 100 x 1.5 mL. Sample vials with pierceable cap 8. Whole Blood Primer, 10 x 1.0 mL (3), Lypholized human red blood cells hemolysate with preservative; 9. Instruction Manual Bio-Rad Laboratories, Hercules, CA 94547. Percentage determination of hemoglobin A1c in human whole blood using ion-exchange high=performance liquid chromatography (HPLC)

Recall number
Z-0023-2012
Recalling firm
Bio-Rad Laboratories, Inc.
Product
VARIANT B-thalassemia Short Program Reorder Pack, 250 Tests, For the separation and determination os HbA and HbF and as an aid in the identification of abnormal hemoglobins in whole blood. IVD, Made in United States. Contents: 1. Elution Buffer 1, 2 x 1900 mL sodium phosphate buffer 2. Elution Buffer 2, 1 x 1800 mL sodium phosphate buffer 3. Wash/Diluent solution, 1 x 1800 mL deionized water 4. Analytical cartridge, 1 cation exchange cartridge 5. 1 Calibrator/Diluent set, 1 Lypholized human red blood cells hemolysate with preservative; deionized water with preservative 6. 1 CD Rom - CD R with program parameters 7 Sample vials, 100 x 1.5 mL. Sample vials with pierceable cap 8. Whole Blood Primer, 10 x 1.0 mL (3), Lypholized human red blood cells hemolysate with preservative; 9. Instruction Manual Bio-Rad Laboratories, Hercules, CA 94547. Percentage determination of hemoglobin A1c in human whole blood using ion-exchange high=performance liquid chromatography (HPLC)
Status
Terminated
Initiated
2011-09-12
Posted
2011-10-07
Terminated
2012-05-11
Reason
A previous recall related to reports of calibration failure and late retention times using Elution Buffer led to an examination of other buffers manufactured during the same time period and a decision was made to field correct these products until a definitive root cause and corrective action have been found and implemented.
Root cause
Process control
Action
Bio-Rad sent an Urgent Medical Device Correction letter on September 12, 2011, to all affected subsidiaries and customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discard all affected product in accordance with all local, state and federal disposal procedures. If customers distributed the product further they should direct their customers to discard all unused product. Customers were asked to please translate (if necessary) the attached "Customer Medical Device Correction Response Form," customers should add their local contact information and distribute to all affected customers in their region by mail or fax. Compile tracking documentation (e.g. Excel spreadsheet) for all customers who received the Medical Device Correction Letter. Complete the "SUBSIDIARY MEDICAL DEVICE CORRECTION RESPONSE FORM" and fax to Bio-Rad Regulatory Affairs at (510) 741-3954, or e-mail PDF to CSD_RA/Hercules/US/BIO-RAD. Customers would be issued credits for the affected products based on their request. For questions call (510) 741-3954.
Quantity
22
Distribution
Nationwide and to New Zealand, Canada, France, England, China, India, Mexico, Italy, Brazil, Australia, Singapore, Thailand, SriLanka, hong Kong, Sweden, Germany, Peru, Jamaica, Uruguay, Columbia, Argentina, Chile, Paraguay and Korea.
Lot, serial or model codes
Model number: 270-2103; Lot number 70210959, expiry 6/30/12. Contains Buffer Lot number AA10685/AA10686, expiry 3/31/11.
510(k) numbers
["K991127"]
PMA numbers
not on file
Product code
LCP
Firm FEI
2915274
Firm city
Hercules
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register