Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Neuromodulation: Medtronic SynchroMed EL Programmable Pumps, Models 8627-10, 8627-18, 8627L-10, 8627L-18. The implantable Medtronic SynchroMed EL Programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories.

Recall number
Z-0022-2007
Recalling firm
Medtronic Neuromodulation
Product
Medtronic SynchroMed EL Programmable Pumps, Models 8627-10, 8627-18, 8627L-10, 8627L-18. The implantable Medtronic SynchroMed EL Programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site. The implantable components of the SynchroMed EL Infusion System include the pump with or without a side catheter access port, catheters, and catheter accessories.
Status
Terminated
Initiated
2006-06-06
Posted
2006-10-20
Terminated
2010-07-06
Reason
The Catheter Access Port (CAP) on SynchroMed EL pumps manufactured between March and July 1999 may detach from the main body of the pump, which can interrupt drug flow to the target site.
Root cause
Process design
Action
An Urgent medical Device Recall letter was sent to healthcare providers beginning June 06, 2006. The letter describes the issue, patient risk, identifies affected devices, and provides patient Management recommendations. The letter also asks for a response card to be returned to Medtronic. Patients may contact Medtronic Neurological Patient Services at 1-800-510-6735.
Quantity
4,317 pumps
Distribution
Worldwide distribution ---- including USA, Argentina, Barbados, Canada, Germany, India, Ireland, Italy, Mexico, The Netherlands, Spain, Switzerland, and United Kingdom.
Lot, serial or model codes
Model 8627-10, serial numbers: NGE000001 - NGE000242R .  Model 8627-18, serial numbers: NGF000001 - NGF003266R.  Model 8627L-10, serial numbers: NGG000001R - NGG000068R,  Model 8627L-18, serial numbers: NGH000001R - NGH001780R.
510(k) numbers
not on file
PMA numbers
["P860004S042"]
Product code
LKK
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register