Skip to the content

FDA Medical Device Recalls (openFDA)

Galemed Corporation: Babi.Plus Bubble PAP Valve with 12.5 cm H2O Pressure Relief Manifold, REF 1002, Rx Only, UDI: 24710810093559

Recall number
Z-0021-2020
Recalling firm
Galemed Corporation
Product
Babi.Plus Bubble PAP Valve with 12.5 cm H2O Pressure Relief Manifold, REF 1002, Rx Only, UDI: 24710810093559
Status
Terminated
Initiated
2019-05-30
Posted
2019-12-20
Terminated
2024-05-20
Reason
Reports of pressure relief manifold venting gas below the standard pressure.
Root cause
Device Design
Action
On 05/30/2019, Medical Device Recall notices were sent to customers via email instructing them to do the following: Within 1 business day, acknowledge receipt of this recall notice. As soon as possible, within 5 business days, after this notification, notify your customers: a) That a product recall been initiated on the affected products. b) Indicate the nature of the recall. 1. Request that the affected lot codes be returned. 2. Customers who do not or cannot return affected product should be cautioned of the events which has led to this recall. 3. Within 5 business days of this notice, provide a list of all customers contacted with product code and lot information. The recalling firm will provide information on processing of affected product. In addition, they will provide an estimated date when properly validated product will be available for shipment. Customers with additional questions were encouraged to contact the firm at sales@galemed.com. Another recall notice, dated 09/19/19, was sent to customers and posted on the recalling firm's website (http://www.galemed.com/). Customers were asked to return reworked product.
Quantity
12000
Distribution
US distribution to distributors in the states of: NC and NV.
Lot, serial or model codes
UDI: 24710810093559 Lot: 180718, 181107, 181127, 190220, and 190220-190610 (Reworked Product)
510(k) numbers
["K110383"]
PMA numbers
not on file
Product code
BZD
Firm FEI
3003319165
Firm city
I-Lan
Firm state
not on file
Firm country
Taiwan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.