Skip to the content

FDA Medical Device Recalls (openFDA)

OriGen Biomedical, Inc.: The OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of the device. A thin sheath is used to cover and secure the wire layer and is heat shrunk into a permanent position.

Recall number
Z-0021-2018
Recalling firm
OriGen Biomedical, Inc.
Product
The OriGen Reinforced Dual Lumen catheter (RDLC) is a duel lumen catheter with a wire reinforcement layer on the tip of the device. A thin sheath is used to cover and secure the wire layer and is heat shrunk into a permanent position.
Status
Terminated
Initiated
2017-08-02
Posted
2017-10-20
Terminated
2024-02-29
Reason
Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss).
Root cause
Under Investigation by firm
Action
The firm initiated their recall by letter on 08/02/2017, following with a press release on 08/28/2017. The firm sent a revised letter to include instructions for the distributors on 09/22/2017 and an updated press release was sent on 09/26/2017. The firm requested that the consignee discontinue use of all OriGen VV28F Reinforced Dual Lumen ECMO Catheters from lots N18487 and N18487-1 and return any of those lots to OriGen Biomedical immediately for replacement or product credit.
Quantity
180 units
Distribution
US Nationwide and the countries of Austria, Belgium, Chile, Czech Republic, Italy, Kingdom of Saudi Arabia, Netherlands, Poland, Sweden, Thailand, Colombia, India, South Africa, United Arab Emirates.
Lot, serial or model codes
Lot Numbers: N18487, N18487-1
510(k) numbers
["K113869"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1000160256
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register