FDA Medical Device Recalls (openFDA)
Accutron Inc: Ultra PC% Cabinet Mount Package G, Model #36600, Serial # 13593, 13594, 13595, 13596, 13597, 13598, 13599, 13600, 13602, 13605, 13610, 13611, 13612, 13615, 13616, 13617, 13621, 13622, 13624, 13625, 13627, 13628, 13633, 13634, 13635, 13636, 13637, 13638, 13639, 13640, 13642, 13643, 13644, 13647, 13650, 13669, 13675, 13673, 13678, 13679, 13680, 13683, 13685, 13687, 13691, 13693, 13694, 13695, 13696, 13697, 13701, 13702, 13705, 13706, 13709, 13710, 13711, 13724, 13725, 13728, 13729, 13735, 13739, 13743, 13746, 13750, 13751, 13753, 13754, 13755, 13756, 13760, 13762, 13766, 13774, 13775, 13777, 13781, 13782, 13783, 13785, 13786, 13788, 13792, 13794, 13796, 13799, 13800, 13803, 13804, 13809, 13810, 28298, 28299, 28304, 28305, 28306, 28307, Ultra PC% Cabinet Mount Pkg I, Model #36800, Serial # 13601, 13603, 13607, 13608, 13609, 13613, 13618, 13620, 13623, 13626, 13629, 13630, 13631, 13632, 13641, 13645, 13646, 13648, 13649, 13651, 13652, 13653, 13654, 13655, 13656, 13657, 13658, 13659, 13660, 13661, 13665, 13666, 13667, 13668, 13671, 13672, 13674, 13676, 13677, 13684, 13688, 13689, 13698, 13700, 13707, 13723, 13727, 13730, 13732, 13734, 13741, 13742, 13757, 13758, 13759, 13761, 13762, 13763, 13764, 13765, 13767, 13769, 13789, 13790, 13791, 13795, 13797, 13801, 13802, 13806, 13808, 28300, 28302, 28303, Ultra PC% Cabinet Mount Pkg G, Model#36600-NS, Serial# 13805, 13807, 13811, 13812. Ultra PC% Cabinet Mount Pkg H, Model#36800-NS, Serial# 13744, 13745,13752, 13773, 13774, 13778, 13779, 13780 Product Usage: Conscious Sedation Unit Analgesia Gas Machine.
- Recall number
- Z-0018-2013
- Recalling firm
- Accutron Inc
- Product
- Ultra PC% Cabinet Mount Package G, Model #36600, Serial # 13593, 13594, 13595, 13596, 13597, 13598, 13599, 13600, 13602, 13605, 13610, 13611, 13612, 13615, 13616, 13617, 13621, 13622, 13624, 13625, 13627, 13628, 13633, 13634, 13635, 13636, 13637, 13638, 13639, 13640, 13642, 13643, 13644, 13647, 13650, 13669, 13675, 13673, 13678, 13679, 13680, 13683, 13685, 13687, 13691, 13693, 13694, 13695, 13696, 13697, 13701, 13702, 13705, 13706, 13709, 13710, 13711, 13724, 13725, 13728, 13729, 13735, 13739, 13743, 13746, 13750, 13751, 13753, 13754, 13755, 13756, 13760, 13762, 13766, 13774, 13775, 13777, 13781, 13782, 13783, 13785, 13786, 13788, 13792, 13794, 13796, 13799, 13800, 13803, 13804, 13809, 13810, 28298, 28299, 28304, 28305, 28306, 28307, Ultra PC% Cabinet Mount Pkg I, Model #36800, Serial # 13601, 13603, 13607, 13608, 13609, 13613, 13618, 13620, 13623, 13626, 13629, 13630, 13631, 13632, 13641, 13645, 13646, 13648, 13649, 13651, 13652, 13653, 13654, 13655, 13656, 13657, 13658, 13659, 13660, 13661, 13665, 13666, 13667, 13668, 13671, 13672, 13674, 13676, 13677, 13684, 13688, 13689, 13698, 13700, 13707, 13723, 13727, 13730, 13732, 13734, 13741, 13742, 13757, 13758, 13759, 13761, 13762, 13763, 13764, 13765, 13767, 13769, 13789, 13790, 13791, 13795, 13797, 13801, 13802, 13806, 13808, 28300, 28302, 28303, Ultra PC% Cabinet Mount Pkg G, Model#36600-NS, Serial# 13805, 13807, 13811, 13812. Ultra PC% Cabinet Mount Pkg H, Model#36800-NS, Serial# 13744, 13745,13752, 13773, 13774, 13778, 13779, 13780 Product Usage: Conscious Sedation Unit Analgesia Gas Machine.
- Status
- Terminated
- Initiated
- 2012-04-29
- Posted
- 2012-10-09
- Terminated
- 2013-02-01
- Reason
- Accutron is recalling the Flowmeter (Analgesia Gas Machine) because when it flows Nitrous Gas without Oxygen gas being turned on, it functions in an incorrect manner.
- Root cause
- Device Design
- Action
- Accutron sent an Urgent Advisory Notice letter dated April 6, 2012 to all their customers who purchased the Ultra PC% Cabinet Mount Flowmeter. The letter informed the customers of the problem identified and actions to be taken. Customers were instructed to discontinue use or to continue use of the device only with extreme care until a replacement can be installed. Extreme care includes the use of the titration method for nitrous oxide administration coupled with close monitoring of both the patient and the actual gas flow through the flowtubes. For question contact Accutron at (800) 531-2221 or by email at jheadrick@accutron-inc.com.
- Quantity
- 220 units total
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- Model #36600, Lot # 73778, 73740, 73851, 73920, 74475, 74035, 74012, 74219, 74118, 74471, 75583, 74372, 75049, 74745, 75049, 75283, 75616, 75299, 74926, 75767, 75344, 75311, 74513. Model #36800, Lot# 73739, 73852, 74011, 73921, 74470, 74032, 74011, 73969, 74139, 74125, 74204, 74220, 74143, 74617, 74573, 74746, 75584, 75449, 74927, 75617, 74568 Model#36600-NS, Lot#75328. Model#36800-NS, Lot#75286.
- 510(k) numbers
- ["K052335"]
- PMA numbers
- not on file
- Product code
- BSZ
- Firm FEI
- 2020813
- Firm city
- Phoenix
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28