FDA Medical Device Recalls (openFDA)
Panoramic Rental Corp.: Panoramic X-ray Model PC-1000, Laser-1000 packed in a crate Product Usage: The PC-1000 will enable the user to take panoramic x-ray images. The PC-1000/Laser1000 will enable the user to take panoramic x-ray images as well as cephalometric x-ray images.
- Recall number
- Z-0016-2017
- Recalling firm
- Panoramic Rental Corp.
- Product
- Panoramic X-ray Model PC-1000, Laser-1000 packed in a crate Product Usage: The PC-1000 will enable the user to take panoramic x-ray images. The PC-1000/Laser1000 will enable the user to take panoramic x-ray images as well as cephalometric x-ray images.
- Status
- Terminated
- Initiated
- 2016-08-19
- Posted
- not on file
- Terminated
- 2018-01-11
- Reason
- Panoramic Corporation is performing a voluntary Medical Device Correction on certain PC-1000 X-ray systems due to the potential for a safety nut to be ineffective, which could result in blunt force trauma that may impact the head, neck or shoulders.
- Root cause
- Device Design
- Action
- Panoramic Corporation sent an Urgent Medical Device Correction letter dated August 19, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached form and return to Panoramic. For questions contact Panoramic at correction@pancorp.com or 888-381-2577 Monday through Thursday 8:00 AM to 6:00 PM and Friday 8:00 AM to 5:00 PM EST in order to verify the effectiveness of the safety component.
- Quantity
- 13,340
- Distribution
- US Nationwide Distribution including Puerto Rico
- Lot, serial or model codes
- Model Number(s): 800724-1, 800724-3, Serial numbers 6001 to 19340
- 510(k) numbers
- ["K870236","K882436"]
- PMA numbers
- not on file
- Product code
- EHD
- Firm FEI
- 1832462
- Firm city
- Fort Wayne
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.