Skip to the content

FDA Medical Device Recalls (openFDA)

Panoramic Rental Corp.: Panoramic X-ray Model PC-1000, Laser-1000 packed in a crate Product Usage: The PC-1000 will enable the user to take panoramic x-ray images. The PC-1000/Laser1000 will enable the user to take panoramic x-ray images as well as cephalometric x-ray images.

Recall number
Z-0016-2017
Recalling firm
Panoramic Rental Corp.
Product
Panoramic X-ray Model PC-1000, Laser-1000 packed in a crate Product Usage: The PC-1000 will enable the user to take panoramic x-ray images. The PC-1000/Laser1000 will enable the user to take panoramic x-ray images as well as cephalometric x-ray images.
Status
Terminated
Initiated
2016-08-19
Posted
not on file
Terminated
2018-01-11
Reason
Panoramic Corporation is performing a voluntary Medical Device Correction on certain PC-1000 X-ray systems due to the potential for a safety nut to be ineffective, which could result in blunt force trauma that may impact the head, neck or shoulders.
Root cause
Device Design
Action
Panoramic Corporation sent an Urgent Medical Device Correction letter dated August 19, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete the attached form and return to Panoramic. For questions contact Panoramic at correction@pancorp.com or 888-381-2577 Monday through Thursday 8:00 AM to 6:00 PM and Friday 8:00 AM to 5:00 PM EST in order to verify the effectiveness of the safety component.
Quantity
13,340
Distribution
US Nationwide Distribution including Puerto Rico
Lot, serial or model codes
Model Number(s): 800724-1, 800724-3, Serial numbers 6001 to 19340
510(k) numbers
["K870236","K882436"]
PMA numbers
not on file
Product code
EHD
Firm FEI
1832462
Firm city
Fort Wayne
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register