Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc: LifeCare PCA Plus II, list #01950-04-03, 01950-04-05, 01950-04-07, 01950-04-51; AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.

Recall number
Z-0016-2010
Recalling firm
Hospira Inc
Product
LifeCare PCA Plus II, list #01950-04-03, 01950-04-05, 01950-04-07, 01950-04-51; AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.
Status
Terminated
Initiated
2009-08-11
Posted
2009-10-06
Terminated
2010-11-30
Reason
Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Root cause
Component design/selection
Action
Firm initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and an Urgent: Device Field Correction letter to consignees was sent via UPS. The letter recommended that customers check their infusion pumps and spare parts in their inventory to determine if the specified AC Power Cords are in use at their facility; and, to inspect the affected cords for bent or cracked prongs and bridge connection, burnt plastic, or excessive wear and tear. Also, it was recommended that customers immediately discontinue use of the affected cords exhibiting the described characteristics and contact their local Hospira representative for assistance. The letter stated that customers may continue to use and monitor cords that are not exhibiting the described characteristics until replacement cords are sent to their facility. Hospira will contact customers to schedule replacement. Customers are to complete and return the attached Reply Form. Question should be directed to Hospira Customer Service at 1-800-241-4002.
Quantity
204,999 total
Distribution
Worldwide Distribution
Lot, serial or model codes
AC power cords Part Numbers: 826-85559-001,  826-85559-003,  826-85559-008,  826-85559-010, and  826-96057-003.
510(k) numbers
["K912928"]
PMA numbers
not on file
Product code
FRN
Firm FEI
2921482
Firm city
Morgan Hill
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register