FDA Medical Device Recalls (openFDA)
Philips Electronics North America Corporation: Philips Healthcare Brilliance 64 System, X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
- Recall number
- Z-0015-2018
- Recalling firm
- Philips Electronics North America Corporation
- Product
- Philips Healthcare Brilliance 64 System, X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
- Status
- Terminated
- Initiated
- 2017-01-18
- Posted
- not on file
- Terminated
- 2020-02-26
- Reason
- During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
- Root cause
- Process design
- Action
- Philips will bring defect into compliance: 1. Will issue a customer notification letter. 2. Will provide a software update, at no cost to the customer. CDRH approves the CAP subject to the following conditions: 1.Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. You will implement this CAP by April 30, 2018. If you need any further information or support concerning this issue, please contact your local Philips representative or the Customer Care Solutions Center at 1-800-722 9377. .
- Quantity
- US - 606
- Distribution
- USA (nationwide) Distribution
- Lot, serial or model codes
- Software version 4.16
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28