FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: Toshiba Aquilion CT System TSX-301A

Recall number
Z-0014-2016
Recalling firm
Toshiba American Medical Systems Inc
Product
Toshiba Aquilion CT System TSX-301A
Status
Terminated
Initiated
2015-05-08
Posted
2015-10-09
Terminated
2016-10-26
Reason
If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved.
Root cause
Radiation Control for Health and Safety Act
Action
Toshiba America Medical Planned Action to bring defect into Compliance: a. A notification will be sent to all consignees with affected systems in compliance with 21 CFR1 003.21 upon approval by the Secretary. b. To correct this issue revised software will be installed on each system to prevent this occurrence from happening. Each consignee will be contacted by their Toshiba service representative for an appointment when the revised software becomes available. c. Each consignee will be advised to do the following until the corrective measures have been taken, i.e., revised software installed on their systems. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Toshiba America Medical Systems, Inc. will implement this CAP by February 1, 2016. If you have any questions please contact your local Toshiba Representative at (800) 521-1968.
Quantity
459
Distribution
US Distribution including Puerto Rico
Lot, serial or model codes
system software V6.0 or later
510(k) numbers
["K113466"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register