Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: Centricity Universal Viewer 7.0, Model 5829628-0XX. Radiological image processing system

Recall number
Z-0012-2022
Recalling firm
GE Healthcare, LLC
Product
Centricity Universal Viewer 7.0, Model 5829628-0XX. Radiological image processing system
Status
Terminated
Initiated
2021-08-12
Posted
2021-10-02
Terminated
2024-08-22
Reason
A software anomaly exists in the Centricity Universal Viewer study management feature in which study changes are not propagated to either the Centricity Enterprise Archive (EA) or another Vendor Neutral Archive (VNA).
Root cause
Other
Action
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected software versions on August 13, 2021. This letter informs customers of the issues and provides safety instructions. GE Healthcare will correct the issue through a product change. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
49
Distribution
AR, IL, ME, MS, NE, NJ, NY, and TN Belgium, Canada, Costa Rica, France, Germany, India, Japan, Korea, Kuwait, Saudi Arabia, Singapore, Switzerland, Switzerland, Turkey and United Kingdom
Lot, serial or model codes
Centricity Universal Viewer, Software Versions 7.0 SP0.0.4.5 and 7.0 SP0.0.5 GTIN 00840682145794
510(k) numbers
["K182419"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register