FDA Medical Device Recalls (openFDA)
Philips Medical Systems North America Co. Phillips: Essenta DR is a digital multifunctional X-ray systems, suitable for all routine radiographic exams, including specialist areas like trauma or pediatric work, excluding mammography. It is designed for radiographic examination of the standing or seated patient or the recumbent patient in combination with a mobile X-ray table (trolley). The intended use is from direct digital imaging using the built in flat panel detector and in addition for free exposures on radiographic cassettes.
- Recall number
- Z-0012-2011
- Recalling firm
- Philips Medical Systems North America Co. Phillips
- Product
- Essenta DR is a digital multifunctional X-ray systems, suitable for all routine radiographic exams, including specialist areas like trauma or pediatric work, excluding mammography. It is designed for radiographic examination of the standing or seated patient or the recumbent patient in combination with a mobile X-ray table (trolley). The intended use is from direct digital imaging using the built in flat panel detector and in addition for free exposures on radiographic cassettes.
- Status
- Terminated
- Initiated
- 2008-08-15
- Posted
- 2010-10-06
- Terminated
- 2010-10-07
- Reason
- When a patient is selected in the patient list and examination tab pressed with background query running at the same time, the wrong patient may be selected.
- Root cause
- Other
- Action
- Philip Medical System sent an "URGENT-Device Correction" letter dated August 13, 2008 to their customers. The letter describes the recalled product as Essenta DR Patient and Image Data Software (SW) release 1.0. The firm informs the customers that when patient is selected in patient list and examination tab pressed with background query running at the same time, there is a risk that the wrong patient is selected on the Essenta DR Computed Radiography System. Customers are advised to verify the selected patient's information before releasing an exposure while the firm works on upgrading the Software. All affected units will receive a software upgrade to correct the issue. Customers should contact Philips Customer Care Service at 800-772-9377, #5, #4, #1 and reference FCO 712000031 for any questions or support concerning the recall.
- Quantity
- 23 units
- Distribution
- Nationwide distribution: CA, GA, FL, KS, MN, NY, OH, OR, SC, TN, TX, UT, and WA.
- Lot, serial or model codes
- Devices were identified as SITE numbers: 553491, 554498, 554944, 553270, 554358, 556089, 556275, 555457, 556080, 557203, 41446107, 41446304, 41443692, 41446097, 41445359, 41445529, 41585399, 41586528, 41796147, 41958550, 42252302, 42616819, and 42865052.
- 510(k) numbers
- ["K070528"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 1000524572
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28