PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Recall number
Z-0010-2026
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Status
Open, Classified
Initiated
2025-09-03
Posted
2025-10-01
Terminated
not on file
Reason
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Root cause
Device Design
Action
On September 3, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. 2. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 4. Complete and return the response form included in this Urgent Medical Device Correction to Philips promptly and no later than 30 days from receipt. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions to be taken by recalling firm: 1. Philips will replace the BIOS battery in all affected systems. 2. Philips expects to start the implementation of this action in affected devices by November 2025. Your local Philips representative will contact you to schedule a visit to perform this activity in your system once available. If you need additional information or support concerning these issues, contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center).
Quantity
300 units (69 US, 231 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Lot, serial or model codes
Model Numbers: 722058; UDI-DI: 00884838059115; Serial Numbers: 337, 178, 249, 22, 202, 279, 254, 285, 71, 186, 82, 85, 162, 138, 24, 365, 203, 167, 146, 168, 304, 172, 200, 123, 129, 149, 259, 4, 197, 134, 38, 290, 171, 56, 264, 238, 199, 61, 156, 99, 65, 59, 109, 100, 305, 319, 215, 278, 124, 196, 405, 325, 265, 378, 277, 274, 94, 182, 366, 340, 190, 219, 373, 239, 23, 205, 257, 212, 111, 110, 399, 72, 332, 92, 142, 67, 106, 335, 360, 415, 11, 114, 151, 93, 232, 395, 292, 380, 183, 349, 261, 15, 375, 161, 78, 368, 9, 372, 288, 252, 320, 213, 293, 389, 169, 97, 336, 25, 143, 133, 192, 49, 412, 201, 89, 66, 163, 170, 77, 273, 35, 222, 229, 248, 334, 384, 40, 268, 240, 258, 289, 269, 226, 244, 48, 141, 391, 241, 164, 381, 281, 154, 246, 127, 315, 256, 233, 348, 204, 396, 7, 185, 95, 131, 377, 60, 68, 206, 101, 150, 333, 84, 398, 313, 3, 39, 344, 376, 382, 370, 387, 419, 394, 407, 413, 392, 411, 358, 409, 385, 416, 367, 379, 331, 418, 356, 421, 369, 420, 397, 357, 408, 393, 388, 108, 211, 126, 125, 301, 12, 30, 355, 354, 159, 139, 128, 81, 58, 96, 343, 177, 115, 19, 18, 311, 220, 386, 291, 327, 214, 145, 223, 137, 136, 119, 103, 404, 27, 32, 350, 400, 422, 423, 347, 73, 43, 55, 194, 189, 237, 230, 221, 247, 116, 242, 29, 37, 276, 345, 166, 179, 53, 87, 303, 90, 112, 245, 148, 41, 234, 208, 117, 193, 76, 236, 266, 130, 235, 251, 86, 31, 195, 144, 330, 63, 295, 70, 187, 79, 10, 33, 75, 361, 153, 284, 102, 21, 64, 45, 36, 181, 351, 272, 210, 243, 20, 98, 34, 16, 218;
510(k) numbers
["K041949","K130842"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register