Skip to the content

FDA Medical Device Recalls (openFDA)

Optovue, Inc.: iVue 500 with iScan, D216606. Product non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber of the eye. The iVue 500 with normative database is a quantitative tool for the comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects. The iVue 500 with normative database is indicated for use as a device to aid in the diagnosis, documentation, and management of ocular health and diseases in the adult population.

Recall number
Z-0010-2018
Recalling firm
Optovue, Inc.
Product
iVue 500 with iScan, D216606. Product non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber of the eye. The iVue 500 with normative database is a quantitative tool for the comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects. The iVue 500 with normative database is indicated for use as a device to aid in the diagnosis, documentation, and management of ocular health and diseases in the adult population.
Status
Terminated
Initiated
2017-07-27
Posted
not on file
Terminated
2021-04-12
Reason
FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).
Root cause
Other
Action
Optovue sent an Urgent Medical Device Software Field Correction letter dated July 27, 2017, via FedEx mail. Customers were Instructed to discontinue use of the Vault Mapping software, to complete and return the recall response form, and to provide remote access to units to an Optovue representative so that new software can be downloaded remotely. Questions can be addressed by calling Optovue at 1-866-941-9240.
Quantity
326 devices
Distribution
US Nationwide Distribution in the states of - AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, & WV.
Lot, serial or model codes
Cat No. 700-49428-001 (GTIN - N/A), 700-49428-002 (GTIN 00858848006315), 700-49428-003 (GTIN 00858848006018).
510(k) numbers
["K133892"]
PMA numbers
not on file
Product code
HLI
Firm FEI
3005950902
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register