Skip to the content

FDA Medical Device Recalls (openFDA)

Hospira Inc.: Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software Product Usage: The Plum A+3 is a dual-line triple channel volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With their primary line, secondary line, and piggyback fluid delivery capability, the Plum A+3 is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories and the LifeShield needleless protection systems makes the Plum A+3 a convenient and cost-effective infusion system.

Recall number
Z-0010-2012
Recalling firm
Hospira Inc.
Product
Plum A+3 Triple Channel Infusion Pumps; Hospira, Inc., Lake Forest, IL 60045; the pumps were sold under the following configurations: a) list 12348 - software version 10.3; b) list 12618 - software version 11.3; c) list 20678 - Hospira MedNet software Product Usage: The Plum A+3 is a dual-line triple channel volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With their primary line, secondary line, and piggyback fluid delivery capability, the Plum A+3 is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories and the LifeShield needleless protection systems makes the Plum A+3 a convenient and cost-effective infusion system.
Status
Terminated
Initiated
2011-09-16
Posted
2011-10-04
Terminated
2016-11-07
Reason
Hospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This condition could cause a delay in therapy if the clinician is unable to start the device due to the continuous recycling and/or rebooting.
Root cause
Software design
Action
The firm, Hospira sent "URGENT DEVICE RECALL" letters dated September 16, 2011 to all affected customers. The letters identified the products, problem and action(s) to be taken. The customers were instructed to read the information provided carefully, complete the attached Reply Form and return it to Hospira via fax at 1-888-345-5358. The letter states that a software upgrade to address the issue is currently being developed. Customers were instructed to notify their consignees if they have further distributed the affected product. The customers were requested to contact Hospira Advanced Knowledge Center at 1-800-241-4002, option 4, for technical assistance. *** Follow-up letters dated June 26, 2012 were sent to the accounts listing the device modules with software versions v13.2, v13.3, v13.40 and v13.5 that will require the software upgrade. ***
Quantity
24,992 pumps
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and the US Virgin Islands, and countries of: Argentina, Australia, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United kingdom, United Arab Emirates, Uruguay and Vietnam.
Lot, serial or model codes
a) list 12348 - serial numbers 0013740398 through 0013741025; b) list 12618 - serial numbers 0014040007 through 0018857420; c) list 20678 - serial numbers 0015543071 through 0019655471
510(k) numbers
["K031185","K042081","K052052"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3013319212
Firm city
Lake Forest
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register