Skip to the content

FDA Medical Device Recalls (openFDA)

Biosite Inc: Multi-Drug 8 Panel Integrated iCup Drug Screen Test Cup. Part Number DUE-187-071-08A. Subsequent Product Codes: DIO, DIS, DJC, DJG, DJR, JXM, JXN, LCM, LDJ, LFG

Recall number
Z-0010-2010
Recalling firm
Biosite Inc
Product
Multi-Drug 8 Panel Integrated iCup Drug Screen Test Cup. Part Number DUE-187-071-08A. Subsequent Product Codes: DIO, DIS, DJC, DJG, DJR, JXM, JXN, LCM, LDJ, LFG
Status
Terminated
Initiated
2009-04-27
Posted
2009-10-06
Terminated
2012-04-02
Reason
The cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.
Root cause
Process control
Action
Biosite, Inc. issued a "Notice of Correction" letter dated May 1, 2009 to consignees describing the problem and instructed users of proper use in order to avoid any cross contamination of samples until all affected product is completely used. For further information, contact Biosite, Inc. using the following information: Biosite Incorporated, an Inverness Medical Company 9975 Summers Ridge Road San Diego, CA 92121 Phone: 1-877-441-7440, option 2 Fax: 1-858-695-7100 Email: productadvisory@invmed.com
Quantity
113,575 cups distributed in total
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot Numbers: DOA8060786, DOA8060787 and DOA8090822.
510(k) numbers
["K061718"]
PMA numbers
not on file
Product code
DKZ
Firm FEI
1000125596
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register