FDA Medical Device Recalls (openFDA)
Optovue, Inc.: iVue with Normative Database, Device Listing Number D185549. Product Usage: The iVue (K121739) is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional, and three dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber of the eye. The iVue with Normative Database is a quantitative tool for the comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects. The iVue with Normative Database is indicated for use as a device to aid in the diagnosis, documentation, and management of ocular health and diseases in the adult population.
- Recall number
- Z-0008-2018
- Recalling firm
- Optovue, Inc.
- Product
- iVue with Normative Database, Device Listing Number D185549. Product Usage: The iVue (K121739) is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional, and three dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber of the eye. The iVue with Normative Database is a quantitative tool for the comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects. The iVue with Normative Database is indicated for use as a device to aid in the diagnosis, documentation, and management of ocular health and diseases in the adult population.
- Status
- Terminated
- Initiated
- 2017-07-27
- Posted
- not on file
- Terminated
- 2021-04-12
- Reason
- FDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).
- Root cause
- Other
- Action
- Optovue sent an Urgent Medical Device Software Field Correction letter dated July 27, 2017, via FedEx mail. Customers were Instructed to discontinue use of the Vault Mapping software, to complete and return the recall response form, and to provide remote access to units to an Optovue representative so that new software can be downloaded remotely. Questions can be addressed by calling Optovue at 1-866-941-9240.
- Quantity
- 326 devices
- Distribution
- US Nationwide Distribution in the states of - AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, & WV.
- Lot, serial or model codes
- Cat No. 700-45575 (GTIN - N/A), 700-45575-002 (GTIN - N/A), 700-45575-A (GTIN - N/A), 700-47818-001 (GTIN - N/A), 700-47818-002 (GTIN - N/A), 700-47818-003 (GTIN - N/A), 700-47818-004 (GTIN 00858848006056), 700-47827-002 (GTIN - N/A), 700-47827-004 (GTIN - N/A), 700-47827-005 (GTIN 00858848006070), 700-49007-002 (GTIN - N/A), 700-49007-003 (GTIN 00858848006216).
- 510(k) numbers
- ["K121739"]
- PMA numbers
- not on file
- Product code
- HLI
- Firm FEI
- 3005950902
- Firm city
- Fremont
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28