Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty

Recall number
Z-0007-2023
Recalling firm
Zimmer, Inc.
Product
G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty
Status
Open, Classified
Initiated
2022-08-25
Posted
2022-10-05
Terminated
not on file
Reason
The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention.
Root cause
Under Investigation by firm
Action
On August 25, 2022, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were instructed to quarantine all affected product. Zimmer Biomet sales representatives will remove the affected product from customer facilities. Customers should ensure that affected personnel are aware of the recall, and also notify any recipients if the product has been further distributed. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Quantity
56 US
Distribution
Domestic distribution US nationwide.
Lot, serial or model codes
Item No. 010000663 UDI-DI: 00880304524224 Lots: 7286477 7289929 7296576 7296633 7301697 Exp. 21-Jul-32 Item No. 010000667 UDI-DI: 00880304524262 Lot: 7253325 Exp. 13-Jun-32
510(k) numbers
["K121874"]
PMA numbers
not on file
Product code
OQI
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register