Abbott Molecular, Inc.: Alinity m SARS-CoV-2 Amp Kit (Non-US Assay) and Alinity m SARS-CoV-2 Application Specification File. in vitro diagnostic

Recall number
Z-0006-2022
Recalling firm
Abbott Molecular, Inc.
Product
Alinity m SARS-CoV-2 Amp Kit (Non-US Assay) and Alinity m SARS-CoV-2 Application Specification File. in vitro diagnostic
Status
Terminated
Initiated
2021-09-02
Posted
2021-10-14
Terminated
2024-07-23
Reason
There is a potential for false positive results.
Root cause
Device Design
Action
The firm issued an Urgent Field Safety Notice/Field Correction Recall 09/02/2021 to all impacted customers. Per the customer letter, all positives are to be considered presumptive until the Alinity m SARS-CoV-2 Application Specification file version 5 or the updated Alinity m Resp-4-Plex Application Specification file is implemented at the customer site. On 03/30/2022, Abbott Molecular issued a follow-up Urgent Field Safety Notice/Field Correction Recall letter to all customers who ordered Alinity m SARS-CoV-2 AMP Kit Emergency Use Authorization (EUA), List Number 09N78-095 and/or Alinity m Resp-4-Plex AMP Kit EUA, List Number 09N79-096 used with the Alinity m System. Also to customers who ordered Alinity m SARS-CoV-2 AMP Kit (CE), List Number 09N78-090 or 09N78-091 and/or Alinity m Resp-4-Plex AMP Kit (CE), List Number 09N79-090 used with the Alinity m System. The Customer Letter explains Abbott has updated the existing Application Specification files for Alinity m SARS-CoV-2 and Resp-4-Plex to further reduce the potential for carryover. FDA has not yet authorized the Alinity m SARS-CoV-2 Application Specification file version 6 update or the Alinity m Resp-4-Plex Application Specification file version 5 update. Implementation of the modification for Alinity m SARS-CoV-2 and Resp-4-Plex will occur when the update to the Application Specification file has been authorized. A Molecular Diagnostics Abbott Representative will be contacting you once the updated Application Specification files are available. If you have any questions, please contact your local Molecular Diagnostics Abbott representative or call Abbott Technical Support at 1-800-553-7042 Option 2.
Quantity
71314 units
Distribution
Worldwide Distribution. US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, COLOMBIA, CZECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, HONG KONG, IRELAND, ISRAEL, ITALY, JAPAN, MALAYSIA, MONTENEGRO, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, UNITED KINGDOM, and VIETNAM.
Lot, serial or model codes
List Number: 09N78-090, UDI UDI 00884999049222; Lot/Serial, Expiration: 511468, 09/15/2021; 511504, 09/15/2021; 511505, 09/15/2021; 512025, 09/21/2021; 512416, 09/18/2021; 512417, 09/18/2021; 512705, 09/21/2021; 512712, 10/07/2021; 512898, 10/07/2021; 513147, 10/14/2021; 513629, 11/18/2021; 514359, 12/01/2021; 514762, 12/17/2021; 515053, 01/05/2022; 515415, 01/15/2022; 515416, 01/15/2022; 515948, 01/28/2022; 516084, 02/05/2022; 516430, 02/17/2022; 516431, 02/18/2022; 516612, 02/25/2022; 516636, 03/01/2022; 516910, 03/04/2022; 516911, 02/11/2022; 517009, 02/18/2022; 517138, 02/25/2022; 517354, 03/08/2022; 517876, 03/19/2022; 517998, 03/19/2022; 518668, 04/01/2022; 518669, 04/20/2022; 518878, 04/20/2022; 519022, 04/30/2022; 519122, 04/30/2022; 519372, 05/03/2022; 519373, 05/11/2022; 519784, 05/18/2022; 519967, 05/25/2022; 520221, 06/08/2022; 520387, 06/15/2022; 520388, 06/15/2022; 520563, 06/18/2022; 527137, 01/20/2023; 527205, 01/20/2023; 527402, 01/28/2023; 527497, 01/28/2023; 527576, 02/04/2023; 527642, 02/08/2023; 527844, 02/17/2023; 528024, 02/18/2023; List Number: 09N78-03E, UDI (01)00884999050211(240)09N78-03E(8012)5.oo
510(k) numbers
not on file
PMA numbers
not on file
Product code
QJR
Firm FEI
3005248192
Firm city
Des Plaines
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register