FDA Medical Device Recalls (openFDA)

ELITech Clinical Systems SAS: ENVOY 500 AST Reagent Kit, reference Model/Catalog 55255, for Envoy 500. Intended Use of the Product: Envoy 500 AST Reagent is for the quantitative in vitro diagnostic determination of aspartate aminotransferase (AST) in human serum and plasma using the Envoy 500 Chemistry System. Aspartate aminotransferase (AST) measurements are used in the diagnosis and treatment of certain types of liver and heart diseases.

Recall number
Z-0006-2015
Recalling firm
ELITech Clinical Systems SAS
Product
ENVOY 500 AST Reagent Kit, reference Model/Catalog 55255, for Envoy 500. Intended Use of the Product: Envoy 500 AST Reagent is for the quantitative in vitro diagnostic determination of aspartate aminotransferase (AST) in human serum and plasma using the Envoy 500 Chemistry System. Aspartate aminotransferase (AST) measurements are used in the diagnosis and treatment of certain types of liver and heart diseases.
Status
Terminated
Initiated
2014-08-11
Posted
2014-10-02
Terminated
2014-12-11
Reason
Some users of ENVOY 500 AST Reagent Kit, reference 55255, for Envoy 500 systems are observing that Quality Control (QC) fails before the product expiry is reached.
Root cause
Process control
Action
ELITech sent an Urgent Product Recall Notice letter dated August 12, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. To All Users, ***Effective immediately, please discontinue using this product.*** Actions to be taken: "Laboratories must not use this product and must destroy all remaining inventory of the product. "If you have transferred any of this product to another location, please forward a copy of this bulletin to that location. "All users receive ENVOY 500 AST REAGENT KIT, reference 55255, from distributors. To be certain everyone who is possibly affected by this is contacted, we are sending this communication to all registered users. As confirmation of receipt of this notice, please return the attached confirmation receipt to Technical Support by fax (401-642-9003) within 48 hours of receipt. On August 20, 2014:, customers were contacted with an interim solution. On September 3, 2014, the firm sent Technical Bulletin 5056 and Distributor Notice D057/Customer Notice C061 to all Envoy users notifying that the recalled Envoy AST Reagent (ref 55255) was being discontinued; the old Envoy AST Reagent (ref. 55250) was being reinstated. On September 23, 2014, the firm sent the URGENT: PRODUCT RECALL NOTICE "UPDATE", dated 19 September 2014, to all Envoy users stating that old (original) Envoy 500 AST Reagent Kit (ref 55250) will replace the recalled Envoy 500 AST Reagent Kit (ref 55255). As confirmation of receipt of this notice, please return the attached confirmation receipt to Technical Support by fax (401-642-9003) within 48 hours of receipt. Customers with questions about this recall, please call the Technical Support at 1-855-354-8324 (1-855-ELITECH).
Quantity
878 kits
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot 4014, Expiration date 07/2015; Lot 4051, Expiration date 09/2015; Lot 4074, Expiration date 11/2015;
510(k) numbers
["K102647"]
PMA numbers
not on file
Product code
CIT
Firm FEI
3007662974
Firm city
SEES
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register