Zyno Medical LLC: Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and 4.1.08z; Z-800W 3.1.32 and 3.1.64z, Z-800WF 3.1.32 and 3.1.64z

Recall number
Z-0005-2025
Recalling firm
Zyno Medical LLC
Product
Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and 4.1.08z; Z-800W 3.1.32 and 3.1.64z, Z-800WF 3.1.32 and 3.1.64z
Status
Open, Classified
Initiated
2024-09-13
Posted
2024-10-09
Terminated
not on file
Reason
There is a defect in the air-in-line software algorithm.
Root cause
Process change control
Action
An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated 9/13/24 was sent to customers. Mitigation Zyno Medical LLC is mitigating this software issue by conducting a voluntary medical device correction to remove the faulty software and correctly install software that is free of this defect. Recommended Actions 1. Actions to Address the Issue During Use Note that only the devices listed in this letter are affected. No other products are involved in this action. Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below. 2. Actions to Correct the Issue Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF Infusion Pump Air-in-Line Software Issue Response Verification Form. This will include checking your inventory for the affected products, verifying serial numbers, confirming product location, reporting any transferred product, and signing and dating the completed form. A representative from Zyno Medical s partner company, Intuvie LLC, will be in contact with you to coordinate instructions on exchanging devices to remediate this action. Pass this notice to the appropriate personnel who need to be informed if the identified devices have been transferred to another location. If you have any questions or would like assistance, please contact your local Business Development Manager. Contact Information: Customerservice@intuvie.com ***Update 1/3/2025***Additional consignees were sent URGENT: CLASS 1 RECALL letters due to additional affected units identified. Recommended Actions 1. Actions to Address the Issue During Use Note that only the devices listed in this letter are affected. No other products are involved in this action. Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below. 2. Actions to Correct the Issue Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF I
Quantity
34,994 units (1819 units still need correction)
Distribution
US Nationwide.
Lot, serial or model codes
UDI-DI: Z-800 - 00814371020006 Z-800W - 00814371020020 Z-800F - 00814371020013 Z-800WF - 00814371020037 Lot Numbers: 190722360 20110921-SH 20111005-SH 20111102-SH 20111212-SH 20120105-SH 20120222-SH 20120501-SH 20120612-SH 20120719-SH 20120809-SH 20190809-SH 20120910-SH-1 20121213-sh 20130123-SH 20130227-SH 20120417-SH 20130511-SH 20130517-SH 20130609-SH 20130701-SH 20130812-SH 20131024-SH
510(k) numbers
["K130690","K100705"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3006575795
Firm city
Natick
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register