FDA Medical Device Recalls (openFDA)
Covidien, LP: Endo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 2.0 mm, Item Code 030453 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.
- Recall number
- Z-0004-2021
- Recalling firm
- Covidien, LP
- Product
- Endo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 2.0 mm, Item Code 030453 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.
- Status
- Open, Classified
- Initiated
- 2020-08-12
- Posted
- not on file
- Terminated
- not on file
- Reason
- The device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
- Root cause
- Under Investigation by firm
- Action
- Urgent Medical Device Recall notification letters dated 8/12/20 were sent to customers. Required Actions: 1.Please immediately quarantine and discontinue use of affected item codes with associated lot numbers listedon Attachment A. 2.Please return affected product as indicated below. All unused products from the affected item codes and associated lot numbers must be returned. 3.If you have distributed the Covidien Endo GIA" Auto Suture" universal articulating loading units listed on Attachment A, please promptly forward the information from this letter to those recipients. 4.Complete the Recalled Product Return Form even if you do not have inventory. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
- Quantity
- 900,007 total
- Distribution
- Worldwide distribution - US Nationwide and Global Distribution.
- Lot, serial or model codes
- Lot Numbers: T7F017X T7J023X T7L032X T7L034X T8L100X T7F025X T7J024X T7L033X T7L035X T9F138X T7F026X T7K020X
- 510(k) numbers
- ["K111825"]
- PMA numbers
- not on file
- Product code
- GDW
- Firm FEI
- 1219930
- Firm city
- North Haven
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28