FDA Medical Device Recalls (openFDA)
Leonhard Lang Medizintechnik GmbH: Skintact Electrodes for Defibrillation, DF29N. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only.
- Recall number
- Z-0004-2017
- Recalling firm
- Leonhard Lang Medizintechnik GmbH
- Product
- Skintact Electrodes for Defibrillation, DF29N. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only.
- Status
- Terminated
- Initiated
- 2016-09-01
- Posted
- 2016-10-14
- Terminated
- 2019-03-26
- Reason
- There is a risk that defibrillation electrodes model DF29N will be connected with the defibrillator Welch Allyn AED 10 only with delay or not at all. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
- Root cause
- Device Design
- Action
- Consignees was sent via e-mail a Leonhard Lang "Important Safety Notice" dated September 1, 2016. The letter described the product being recalled, Description of the defect, Actions and time frame of the recall, & Compensation for the recalled electrodes. Advised consignees to inform all users within their organizations of the recall; and, secure and destroy the unused electrodes in their inventory. Confirmation of destruction can be done by completing and returning the "Confirmation of Destruction / Consumption" form by October 14, 2016. If the product was further distributed, they were to forward a copy of the Safety Notice. For questions contact sales staff.
- Quantity
- 11,110 ( US 8,040; OUS 3,070)
- Distribution
- Worldwide Distribution - US, to the state of Florida; and, the countries of South Africa, Germany, Slovenia, France, United Arab Emirates, Great Britain, Italy, Israel, Serbia, Thailand, Thailand, South Africa, Lebanon, and Poland.
- Lot, serial or model codes
- US: 60602-0774; 60502-0779; 60308-0771; 60114-0773; 51023-0775; 50904-0777; 50403-0778; 50130-0777; 41023-0771; 41008-0778; 40730-0778; 40618-0778; and, 40130-0776. Outside US: 60725-0774; 60620-0776; 60602-0774; 60502-0779; 60308-0771; 60114-0773; 51023-0775; 50904-0777; 50403-0778; 40827-0777; 40730-0778; 40618-0778; and, 40130-0776.
- 510(k) numbers
- ["K142803"]
- PMA numbers
- not on file
- Product code
- MKJ
- Firm FEI
- 6736
- Firm city
- Innsbruck
- Firm state
- not on file
- Firm country
- Austria
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28