Philips Healthcare Inc.: HeartStart MRx (M3535A/M3536A) Defibrillator/Monitor. The HeartStart MRx/MRxE is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

Recall number
Z-0003-2010
Recalling firm
Philips Healthcare Inc.
Product
HeartStart MRx (M3535A/M3536A) Defibrillator/Monitor. The HeartStart MRx/MRxE is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
Status
Terminated
Initiated
2009-08-21
Posted
2009-11-09
Terminated
2017-04-25
Reason
incorrect default configuration settings, insufficient battery power as the user may not be alerted that no battery is installed or that a low battery condition exists prior to disconnecting from AC or DC power
Root cause
Process change control
Action
Philips Healthcare issued a Urgent Voluntary Medical Device Correction Notification on August 21, 2009 advising users of the problem and Customers are asked to follow the "Action to be taken by customer/user" section of the Urgent Voluntary Medical Device Correction Notification. In this section, customers are given the instructions to correct their devices. For further information or support concerning this issue, please contact Philips representative at 1-800-722-9377.
Quantity
1,523 units
Distribution
Nationwide Canada Australia, Austria, Beharin, BRUNEI DARUSSALAM,Chian,. Finland, France, Germany,Hong Kong, India, Indonesia, Ireland, Israel,Italy, Japan, Libya, Lithuania. Mexico, Nambia, Netherlands, Palestine, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, UAE, UK
Lot, serial or model codes
Serial numbers within the following range. US00533518 through US00535118.
510(k) numbers
["K031187"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register