Skip to the content

FDA Medical Device Recalls (openFDA)

Biosense Webster, Inc.: Carto 3 EP Navigation System. Electro physiology system which views of the electrical activity of the heart through real-time data on 3-D, color-coded cardiac maps.

Recall number
Z-0002-2016
Recalling firm
Biosense Webster, Inc.
Product
Carto 3 EP Navigation System. Electro physiology system which views of the electrical activity of the heart through real-time data on 3-D, color-coded cardiac maps.
Status
Terminated
Initiated
2015-09-10
Posted
2015-10-01
Terminated
2016-05-13
Reason
Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.
Root cause
Device Design
Action
A Customer Notification Letter and Acknowledgement Form was distributed to customers on 09/10/15. One letter and form was sent via express mail, addressed to the Risk Management Office and one additional letter and form was sent by the firm's account representative to the Electrophysiology (EP/Cardiology Lab directly at the customer site. The firm's notification letter stated that they want to emphasize the following statement "The intra-cardiac ultrasound image will disappear if the CARTO 3 EP Navigation System power is disrupted and this may present a safety issue if the EP is using the ultrasound to monitor the patien during EP procedure. The ultrasound image will not reappear until the CARTO 3 System is restored." The firm will be updating the product labeling to further reinforce the precautionary statement. The firm requests that customers take the following actions: 1. Read the Field Safety Notification carefully. 2. Pass on this notification to anyone in the facility that needs to be informed of this issue, including appropriate clinical personnel involved in the use of SOUNDSTAR eco Catheters. 3. Review, complete, sign and return the attached Acknowledgement Form in accordance with the instructions on the form. 4. Maintain a copy of this letter with the product. 5. Maintain awareness of the Field Safety Notification For any questions the firm states to contact the BWI sales representative.
Quantity
1035 (U.S.)
Distribution
Worldwide Distribution. US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.
Lot, serial or model codes
Catalog #: C3SOUND, 10439236, 10439072, 10439011, 10438577
510(k) numbers
["K133916"]
PMA numbers
not on file
Product code
DQK
Firm FEI
3003113629
Firm city
Irwindale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register