FDA Medical Device Recalls (openFDA)
Medtronic Navigation, Inc: Probe, Tactile, Right, ASM, Long, Part #962011S, Medtronic Navigation, Louisville, CO 80027. Trackable hand held instruments used in spine surgeries for preparation of the pedicle. These instruments are used in conjunction with the Medtronic Navigation StealthStation platform and various software applications.
- Recall number
- Z-0002-2012
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Probe, Tactile, Right, ASM, Long, Part #962011S, Medtronic Navigation, Louisville, CO 80027. Trackable hand held instruments used in spine surgeries for preparation of the pedicle. These instruments are used in conjunction with the Medtronic Navigation StealthStation platform and various software applications.
- Status
- Terminated
- Initiated
- 2011-01-20
- Posted
- 2011-10-03
- Terminated
- 2011-10-03
- Reason
- Distal end of tactile probe may break off during use.
- Root cause
- Labeling design
- Action
- Medtronic sent an "Urgent Product Safety Notice" to all affected customers dated Jan 20, 2011. The letter included affected product and updated instructions for use. On-site training was given for the use of the probes by Medtronic Representatives. For information on this recall call Medtronics at (720) 890-3409.
- Quantity
- 6 probes
- Distribution
- Worldwide Distribution: Nationwide distribution including the states of CO and WI; and the country of South Korea
- Lot, serial or model codes
- Lot Number 071219
- 510(k) numbers
- ["K990214"]
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 1000517638
- Firm city
- Louisville
- Firm state
- CO
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28