Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc: FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic use.

Recall number
Z-0002-2008
Recalling firm
Beckman Coulter Inc
Product
FP1000 Cell Preparation System Software versions 1.1 & 1.2, Part Number 624922, for in vitro diagnostic use.
Status
Terminated
Initiated
2007-02-22
Posted
2007-10-16
Terminated
2012-04-26
Reason
Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented.
Root cause
Labeling design
Action
A Product Corrective Action (PCA) letter was mailed on Feb 22, 2007, to all FP1000 customers to inform them of an issue associated with the FP1000 Cell Preparation System. Pressure (from recapped tubes) or vacuum (from under-filled short-draw tubes) will cause inaccurate results if not properly vented. Under these conditions the FP1000 will under-dispense specimen volume leading to lower than expected absolute counts. Specimen tubes that are recapped or underfilled must therefore be vented prior to running on the FP1000. If the initial state of the specimen tube is unknown, then it must be vented to ensure accurate results prior to running on the FP1000. Additional instructions are provided. A Fax Back Response Form was also provided.
Quantity
9 in the U.S.; 1 in Canada
Distribution
Nationwide (MA, IN, IL, UT, RI, NJ), and Canada.
Lot, serial or model codes
Part Number 624922. Software versions 1.1 & 1.2
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXG
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register