Terumo Cardiovascular Systems Corporation: CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
- Recall number
- Z-0001-2017
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
- Status
- Terminated
- Initiated
- 2015-12-11
- Posted
- not on file
- Terminated
- 2016-10-26
- Reason
- Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.
- Root cause
- Other
- Action
- Terumo (Planned Action) plans to execute the following steps as part of the correction activity: Notify all customers of the potential that a cuvette may disengage from the circuit (probe) leading to a "disconnect" message on the monitor. This notification will be performed via a Customer Notification Letter. For further questions, please call (410) 392-7277.
- Quantity
- 140,000 units
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution.
- Lot, serial or model codes
- 6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various
- 510(k) numbers
- ["K915265"]
- PMA numbers
- not on file
- Product code
- DRY
- Firm FEI
- 3000204839
- Firm city
- Elkton
- Firm state
- MD
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28