FDA Medical Device Recalls (openFDA)
Penumbra Inc.: Penumbra Neuron Delivery Catheter 070, percutaneous catheter, catalog numbers PND6F0701058, PND6F070956, PND6F0701058M, and PND6F070956M; manufactured by Penumbra, Inc., Alameda, CA. Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
- Recall number
- Z-0001-2010
- Recalling firm
- Penumbra Inc.
- Product
- Penumbra Neuron Delivery Catheter 070, percutaneous catheter, catalog numbers PND6F0701058, PND6F070956, PND6F0701058M, and PND6F070956M; manufactured by Penumbra, Inc., Alameda, CA. Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
- Status
- Terminated
- Initiated
- 2009-09-22
- Posted
- 2009-10-05
- Terminated
- 2010-11-24
- Reason
- Failure-- product can kink or ovalize, which may result in difficulty in catheter advancement or delivery of other devices and damage to catheter, which could lead to patient injury or death as a result of portions being retained in the patient.
- Root cause
- Device Design
- Action
- The firm issued an Urgent Voluntary Field Removal Notice letter on 09/22/2009 to its consignees, asking them to inspect their stocks; remove affected product and place in quarantine for return to Penumbra Inc.; and complete and return the attached product identification / return form.
- Quantity
- 2246 units
- Distribution
- Worldwide Distribution -- USA, Canada, Hong Kong, Chile, the UK, Denmark, Norway, Sweden, Finland, Netherlands, Germany, Switzerland, Austria, and Spain.
- Lot, serial or model codes
- Lot numbers F13733, F13751, F13773, F13799, F13817, F13826, F13856, F13865, F13880, F13899, F13934, F13960, F13970, F13971, F13980, F13997, F14007, F14028, F14045, F14046, F14047, F14048, F14102, F14103, F14150, F14182, F14183, F14427, F14482, F14560, and F14629.
- 510(k) numbers
- ["K082290"]
- PMA numbers
- not on file
- Product code
- DQY
- Firm FEI
- 3005168196
- Firm city
- Alameda
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as