Philips Medical Systems North America Co. Phillips: NeuViz Dual Multi-Slice Computed Tomography (CT) Scanner System (software). Whole body x-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is an interactive program used for x-ray scan control, image reconstruction, and image archive/evaluation. It is intended to produce cross-section images of head and body by computer reconstruction of x-ray transmission data taken at different angles. The function of Time Scan is intended to use in contrast scan include single phase and multi-phase scan, that is according to different density of contrast in vessels or organs between different phase to get necessary information to support doctor's diagnosis.
- Recall number
- Z-0001-2009
- Recalling firm
- Philips Medical Systems North America Co. Phillips
- Product
- NeuViz Dual Multi-Slice Computed Tomography (CT) Scanner System (software). Whole body x-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system software is an interactive program used for x-ray scan control, image reconstruction, and image archive/evaluation. It is intended to produce cross-section images of head and body by computer reconstruction of x-ray transmission data taken at different angles. The function of Time Scan is intended to use in contrast scan include single phase and multi-phase scan, that is according to different density of contrast in vessels or organs between different phase to get necessary information to support doctor's diagnosis.
- Status
- Terminated
- Initiated
- 2008-04-10
- Posted
- 2009-01-26
- Terminated
- 2010-10-29
- Reason
- An issue associated with the "Timed Scan" application was found. The system does not wait for the programmed time interval and starts the x-ray after selection of the scan button when certain criteria is met.
- Root cause
- Software design
- Action
- A letter was issued to customers prior to the release of the field correction during June 2008. The system software will be updated to correct the non-compliances. For information or support concerning this issue, please contact the Neusoft Medical Systems Service Support Department (helpdesk@pnms.neusoft.com) or contact Philips Medical Systems North America Co. Phillips at 425-487-7602.
- Quantity
- 19 units in the US.
- Distribution
- Nationwide Distribution --- USA including states of CO, FL, GA, IL, KY, NC, TN and TX.
- Lot, serial or model codes
- Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH025EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002 and NDHR080003.
- 510(k) numbers
- ["K062451","K071308"]
- PMA numbers
- not on file
- Product code
- JAA
- Firm FEI
- 1000524572
- Firm city
- Bothell
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28